9 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 10, 2016
Product types
Device
Classifications
Class I
Statuses
Terminated
Recalling firm wording
Arrow International, Inc., Division of Teleflex Medical Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

9 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 9

FiberOptix Ultra 8 IAB: 8Fr 30cc The Arrow¿ IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diastole and deflation, during systole increases blood supply to the heart muscle and decreases work of the left ventricle.

Z-1055-2016
Recall number
Z-1055-2016
Initiated
February 10, 2016
Classification
Class I
Status
Terminated
Quantity
13,405 US and 33,735 OUS in total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

Code information

Batch: 18F14A0031 18F14A0032 18F14A0033 18F14D0066 18F14E0018 18F14E0072 18F14F0067 18F14F0079 18F14G0045 18F14G0046 18F14G0081 18F14H0037 18F14J0023 18F14L0014 18F14M0033 18F15A0010 18F15B0015 18F15C0004 18F15C0023 18F15D0025 18F15D0052 18F15E0020 18F15F0002 18F15F0039 18F15G0005 18F15G0015 18F15H0006 18F15H0037 18F15H0051 18F15J0026 18F15L0007 18S14G0046

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Argentina, Australia, Belgium, Brunei Darussalam, Brazil, Bahamas, Canada, Chile, China, Colombia, Ecuador, Guatemala, Hong Kong, Indonesia, India, Japan, Korea (Republic of [South] Korea), Myanmar, Mexico, Malaysia, Nepal, New Zealand, Panama, Peru, Philippines, Pakistan, Singapore, Thailand, Trinidad and Tobago, Taiwan, Uruguay, Viet Nam, and South Africa (Zuid Afrika).

device · product 2 of 9

Ultra 8 IAB: 8Fr 30cc; Product Code: IAB-05830-U The Arrow¿ IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diastole and deflation, during systole increases blood supply to the heart muscle and decreases work of the left ventricle.

Z-1056-2016
Recall number
Z-1056-2016
Initiated
February 10, 2016
Classification
Class I
Status
Terminated
Quantity
13,405 US and 33,735 OUS in total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

Code information

Batch: 18F14A0031 18F14A0032 18F14A0033 18F14D0066 18F14E0018 18F14E0072 18F14F0067 18F14F0079 18F14G0045 18F14G0046 18F14G0081 18F14H0037 18F14J0023 18F14L0014 18F14M0033 18F15A0010 18F15B0015 18F15C0004 18F15C0023 18F15D0025 18F15D0052 18F15E0020 18F15F0002 18F15F0039 18F15G0005 18F15G0015 18F15H0006 18F15H0037 18F15H0051 18F15J0026 18F15L0007 18S14G0046

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Argentina, Australia, Belgium, Brunei Darussalam, Brazil, Bahamas, Canada, Chile, China, Colombia, Ecuador, Guatemala, Hong Kong, Indonesia, India, Japan, Korea (Republic of [South] Korea), Myanmar, Mexico, Malaysia, Nepal, New Zealand, Panama, Peru, Philippines, Pakistan, Singapore, Thailand, Trinidad and Tobago, Taiwan, Uruguay, Viet Nam, and South Africa (Zuid Afrika).

device · product 3 of 9

FiberOptix Ultra 8 IAB: 8Fr 40cc; Product Code: IAB-05840-LWS The Arrow¿ IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diastole and deflation, during systole increases blood supply to the heart muscle and decreases work of the left ventricle.

Z-1057-2016
Recall number
Z-1057-2016
Initiated
February 10, 2016
Classification
Class I
Status
Terminated
Quantity
13,405 US and 33,735 OUS in total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

Code information

Batch: 18F14A0025 18F14A0026 18F14A0030 18F14A0049 18F14A0050 18F14A0051 18F14A0054 18F14A0057 18F14A0066 18F14A0067 18F14D0045 18F14E0004 18F14E0006 18F14E0016 18F14E0017 18F14E0052 18F14F0017 18F14F0047 18F14F0049 18F14F0080 18F14G0003 18F14G0004 18F14G0021 18F14G0037 18F14G0055 18F14G0056 18F14G0068 18F14G0070 18F14G0071 18F14G0076 18F14H0009 18F14H0030 18F14H0054 18F14H0055 18F14H0060 18F14H0061 18F14J0003 18F14J0005 18F14J0037 18F14K0006 18F14M0001 18F14M0014 18F14M0032 18F15A0005 18F15A0011 18F15A0026 18F15A0034 18F15B0005 18F15B0008 18F15B0010 18F15B0018 18F15B0020 18F15B0027 18F15C0002 18F15C0007 18F15C0010 18F15C0015 18F15C0024 18F15C0030 18F15C0036 18F15D0017 18F15D0032 18F15D0034 18F15D0043 18F15D0049 18F15E0003 18F15E0011 18F15E0014 18F15E0023 18F15E0029 18F15E0037 18F15F0011 18F15F0017 18F15F0021 18F15F0032 18F15G0001 18F15G0013 18F15G0021 18F15G0028 18F15G0034 18F15H0010 18F15H0015 18F15H0021 18F15H0024 18F15H0033 18F15J0013 18F15J0028 18F15J0038 18F15J0044 18F15J0049 18F15K0001 18F15K0007 18F15L0002 18S14G0068 18S14M0001 18S15C0002

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Argentina, Australia, Belgium, Brunei Darussalam, Brazil, Bahamas, Canada, Chile, China, Colombia, Ecuador, Guatemala, Hong Kong, Indonesia, India, Japan, Korea (Republic of [South] Korea), Myanmar, Mexico, Malaysia, Nepal, New Zealand, Panama, Peru, Philippines, Pakistan, Singapore, Thailand, Trinidad and Tobago, Taiwan, Uruguay, Viet Nam, and South Africa (Zuid Afrika).

device · product 4 of 9

Ultra 8 IAB: 8Fr 40cc; Product Code: IAB-05840-U The Arrow¿ IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diastole and deflation, during systole increases blood supply to the heart muscle and decreases work of the left ventricle.

Z-1058-2016
Recall number
Z-1058-2016
Initiated
February 10, 2016
Classification
Class I
Status
Terminated
Quantity
13,405 US and 33,735 OUS in total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

Code information

Batch: 18F14B0044 18F14B0045 18F14E0060 18F14F0019 18F14G0067 18F14H0002 18F14J0026 18F15A0033 18F15B0009 18F15B0019 18F15C0012 18F15D0016 18F15D0037 18F15D0044 18F15F0001

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Argentina, Australia, Belgium, Brunei Darussalam, Brazil, Bahamas, Canada, Chile, China, Colombia, Ecuador, Guatemala, Hong Kong, Indonesia, India, Japan, Korea (Republic of [South] Korea), Myanmar, Mexico, Malaysia, Nepal, New Zealand, Panama, Peru, Philippines, Pakistan, Singapore, Thailand, Trinidad and Tobago, Taiwan, Uruguay, Viet Nam, and South Africa (Zuid Afrika).

device · product 5 of 9

UltraFlex IAB: 7.5Fr 30cc; Product Code: IAB-06830-U The Arrow¿ IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diastole and deflation, during systole increases blood supply to the heart muscle and decreases work of the left ventricle.

Z-1059-2016
Recall number
Z-1059-2016
Initiated
February 10, 2016
Classification
Class I
Status
Terminated
Quantity
13,405 US and 33,735 OUS in total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

Code information

Batch: 18F13M0016 18F14A0017 18F14A0018 18F14A0022 18F14A0046 18F14E0054 18F14E0070 18F14G0038 18F14G0078 18F14H0007 18F14J0007 18F14J0025 18F14K0016 18F15A0035 18F15B0006 18F15B0014 18F15C0019 18F15E0028 18F15F0016 18F15G0007 18F15H0009 18F15H0035

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Argentina, Australia, Belgium, Brunei Darussalam, Brazil, Bahamas, Canada, Chile, China, Colombia, Ecuador, Guatemala, Hong Kong, Indonesia, India, Japan, Korea (Republic of [South] Korea), Myanmar, Mexico, Malaysia, Nepal, New Zealand, Panama, Peru, Philippines, Pakistan, Singapore, Thailand, Trinidad and Tobago, Taiwan, Uruguay, Viet Nam, and South Africa (Zuid Afrika).

device · product 6 of 9

UltraFlex IAB: 7.5Fr 40cc; Product Code: IAB-06840-U The Arrow¿ IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diastole and deflation, during systole increases blood supply to the heart muscle and decreases work of the left ventricle.

Z-1060-2016
Recall number
Z-1060-2016
Initiated
February 10, 2016
Classification
Class I
Status
Terminated
Quantity
13,405 US and 33,735 OUS in total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

Code information

Batch:18F13M0017 18F13M0019 18F13M0021 18F14A0012 18F14A0013 18F14A0015 18F14A0016 18F14A0044 18F14A0047 18F14B0042 18F14B0043 18F14D0068 18F14E0005 18F14E0007 18F14E0031 18F14E0053 18F14E0069 18F14F0030 18F14F0031 18F14F0066 18F14G0020 18F14G0039 18F14G0066 18F14H0003 18F14H0010 18F14H0023 18F14H0024 18F14H0056 18F14J0006 18F14J0016 18F14J0027 18F14J0051 18F15A0008 18F15A0015 18F15A0023 18F15B0003 18F15B0007 18F15B0021 18F15C0006 18F15C0011 18F15C0027 18F15D0010 18F15D0019 18F15D0045 18F15E0002 18F15E0012 18F15F0008 18F15F0028 18F15F0038 18F15F0041 18F15G0016 18F15G0027 18F15H0005 18F15H0020 18F15H0032 18F15H0041 18F15J0007 18F15J0010 18F15J0021 18F15J0029 18F15J0039 18F15J0056 18F15K0016 18F15K0032 18F15L0001 18F15L0010

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Argentina, Australia, Belgium, Brunei Darussalam, Brazil, Bahamas, Canada, Chile, China, Colombia, Ecuador, Guatemala, Hong Kong, Indonesia, India, Japan, Korea (Republic of [South] Korea), Myanmar, Mexico, Malaysia, Nepal, New Zealand, Panama, Peru, Philippines, Pakistan, Singapore, Thailand, Trinidad and Tobago, Taiwan, Uruguay, Viet Nam, and South Africa (Zuid Afrika).

device · product 7 of 9

RediGuard IAB: 7Fr 30cc; Product Code: IAB-S730C The Arrow¿ IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diastole and deflation, during systole increases blood supply to the heart muscle and decreases work of the left ventricle.

Z-1061-2016
Recall number
Z-1061-2016
Initiated
February 10, 2016
Classification
Class I
Status
Terminated
Quantity
13,405 US and 33,735 OUS in total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

Code information

Batch:18F14A0041 18F14C0003 18F14C0006 18F14C0021 18F14C0026 18F14C0027 18F14C0028 18F14E0039 18F14E0055 18F14E0058 18F14F0035 18F14F0038 18F14F0054 18F14G0057 18F14G0077 18F14H0011 18F15B0023 18F15C0001 18F15C0008 18F15C0026 18F15D0028 18F15F0027 18F15F0037 18F15G0003 18F15G0011 18F15G0032 18F15H0046 18F15M0009

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Argentina, Australia, Belgium, Brunei Darussalam, Brazil, Bahamas, Canada, Chile, China, Colombia, Ecuador, Guatemala, Hong Kong, Indonesia, India, Japan, Korea (Republic of [South] Korea), Myanmar, Mexico, Malaysia, Nepal, New Zealand, Panama, Peru, Philippines, Pakistan, Singapore, Thailand, Trinidad and Tobago, Taiwan, Uruguay, Viet Nam, and South Africa (Zuid Afrika).

device · product 8 of 9

Percutaneous Insertion Tray; Product Code: IAK-06845 The Arrow¿ IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diastole and deflation, during systole increases blood supply to the heart muscle and decreases work of the left ventricle.

Z-1062-2016
Recall number
Z-1062-2016
Initiated
February 10, 2016
Classification
Class I
Status
Terminated
Quantity
13,405 US and 33,735 OUS in total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

Code information

Batch: 18F14A0037 18F14B0040 18F14E0027 18F14F0027 18F14F0028

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Argentina, Australia, Belgium, Brunei Darussalam, Brazil, Bahamas, Canada, Chile, China, Colombia, Ecuador, Guatemala, Hong Kong, Indonesia, India, Japan, Korea (Republic of [South] Korea), Myanmar, Mexico, Malaysia, Nepal, New Zealand, Panama, Peru, Philippines, Pakistan, Singapore, Thailand, Trinidad and Tobago, Taiwan, Uruguay, Viet Nam, and South Africa (Zuid Afrika).

device · product 9 of 9

Percutaneous Insertion Tray; Product Code: IAK-S7IT The Arrow¿ IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diastole and deflation, during systole increases blood supply to the heart muscle and decreases work of the left ventricle.

Z-1063-2016
Recall number
Z-1063-2016
Initiated
February 10, 2016
Classification
Class I
Status
Terminated
Quantity
13,405 US and 33,735 OUS in total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

Code information

Batch: 18F14A0037 18F14B0040 18F14E0027 18F14F0027 18F14F0028

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Argentina, Australia, Belgium, Brunei Darussalam, Brazil, Bahamas, Canada, Chile, China, Colombia, Ecuador, Guatemala, Hong Kong, Indonesia, India, Japan, Korea (Republic of [South] Korea), Myanmar, Mexico, Malaysia, Nepal, New Zealand, Panama, Peru, Philippines, Pakistan, Singapore, Thailand, Trinidad and Tobago, Taiwan, Uruguay, Viet Nam, and South Africa (Zuid Afrika).