cobas p 512 pre-analytical system Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis.
Due to a false triggering or detection of the lifting gripper READY signal, sample tubes are not correctly placed back in the Rack Tube Transport (RTT) after the decapping process. Therefore, open sample tubes can be dropped in the cobas p 512, spilling the sample material.
These labels are deterministic app interpretations, not FDA categories.
Due to a false triggering or detection of the lifting gripper READY signal, sample tubes are not correctly placed back in the Rack Tube Transport (RTT) after the decapping process. Therefore, open sample tubes can be dropped in the cobas p 512, spilling the sample material.
Code information
Part numbers:05083435001 and 06268854001
Distribution pattern
US Distribution including Puerto Rico and to the states of :TX, OH, TN, AZ, WA, MI and GA