Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 73428

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 18, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Sedecal S.A.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Mobile wDR motorized portable diagnostic X-ray systems. Intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts on both adult and pediatric patients. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not intended for mammography.

Z-1084-2016
Recall number
Z-1084-2016
Initiated
January 18, 2016
Classification
Class II
Status
Terminated
Recalling firm
Sedecal S.A.
Quantity
349 systems

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
This is related to previous 806 Notifications about uncommanded movements of a motorized mobile x-ray system. Reason for Recall: Along with the previously noted encoder and gauges problems and corresponding fixes, we have observed that static electricity can also elicit the unwanted behavior.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

This is related to previous 806 Notifications about uncommanded movements of a motorized mobile x-ray system. Reason for Recall: Along with the previously noted encoder and gauges problems and corresponding fixes, we have observed that static electricity can also elicit the unwanted behavior.

Code information

Serial Numbers: 485886/SN11000007, 485062/SN11000009, 488169/SN11000011, 485061/SN11000012, 485059/SN11000013, 485060/SN11000015, 489089/SN11000016, 489616/SN11000033, 490027/SN11000034, 489694/SN11000035, 489695/SN11000036, 489696/SN11000037, 490221/SN11000039, 489088/SN11000040, 493566/SN11000047, 493565/SN11000048, 494867/SN11000051, 494941/SN11000052, 492417/SN11000053, 494456/SN11000055, 494865/SN11000056, 494864/SN11000060, 495286/SN11000061, 494866/SN11000062, 494464/SN11000065, 497570/SN11000066, 488171/SN11000068, 494060/SN11000070, 494061/SN11000071, 494062/SN11000072, 494063/SN11000073, 494064/SN11000074, 494065/SN11000075, 497715/SN11000080, 496446/SN11000082, 496387/SN12000003, 495938/SN12000007, 495939/SN12000008, 493364/SN12000022, 493462/SN12000023, 501028/SN12000026, 500739/SN12000027, 500740/SN12000028, 502424/SN12000034, 503143/SN12000036, 503142/SN12000037, 503141/SN12000038, 503144/SN12000039, 503541/SN12000040, 500466/SN12000045, 502721/SN12000047, 505077/SN12000048, 506626/SN12000051, 507607/SN12000054, 507906/SN12000055, 508342/SN12000057, 500468/SN12000058, 496401/SN12000059, 500467/SN12000060, 508838/SN12000061, 508532/SN12000062, 504450/SN12000064, 509718/SN12000068, 510607/SN12000072, 512160/SN12000077, 513477/SN12000080, 513478/SN12000081, 512892/SN12000082, 513848/SN12000084, 512887/SN12000085, 512888/SN12000086, 512890/SN12000087, 508699/SN12000088, 514462/SN12000089, 514461/SN12000090, 514104/SN12000091, 514103/SN12000092, 509062/SN12000107, 508783/SN12000108, 516891/SN12000110, 516896/SN12000111, 497572/SN12000112, 518135/SN12000118, 518342/SN12000119, 518343/SN12000120, 518344/SN12000121, 518345/SN12000122, 518346/SN12000123, 502025/SN12000124, 502026/SN12000125, 12000166, 519326/SN13000005, 519388/SN13000006, 519511/SN13000010, 519506/SN13000011, 519387/SN13000012, 519763/SN13000013, 519434/SN13000014, 520251/SN13000016, 520253/SN13000017, 519762/SN13000018, 520254/SN13000019, 520255/SN13000020, 520256/SN13000021, 520413/SN13000023, 520414/SN13000025, 520415/SN13000026, 520252/SN13000027, 520422/SN13000028, 520418/SN13000030, 520416/SN13000031, 520417/SN13000033, 522633/SN13000035, 522944/SN13000037, 523279/SN13000042, 523349/SN13000043, 13000046, 13000047, 13000048, 13000049, 13000053, 13000054, 13000055, 13000056, 13000057, 13000058, SN13000059, 13000060, SN13000061, SN13000062, SN13000063, 13000067, SN13000069, 13000070, 13000071, SN13000072, SN13000073, SN13000074, 13000075, 13000076, 13000077, 13000078, 13000079, SN13000080, SN13000082, SN13000083, SN13000084, 13000085, SN13000087, SN13000089, SN13000090, SN13000091, 13000092, 13000093, 13000094, SN13000099, SN13000100, SN13000101, SN13000102, SN13000103, SN13000104, SN13000106, SN13000107, 13000109, SN13000112, SN13000113, 519598/SN13000118, 519599/SN13000119, 519600/SN13000120, SN13000125, 13000127, 13000139, 13000140, 13000141, 13000143, 13000144, 13000145, SN13000146, SN13000147, SN13000148, SN13000149, SN13000153, SN13000154, SN13000157, SN7369686, SN13000159, SN13000160, SN13000163, SN13000164, SN13000165, 13000167, 13000168, 13000169, SN13000170, SN13000174, SN13000176, SN13000177, SN13000179, SN13000180, SN13000181, SN13000182, SN13000183, SN13000185, 13000196, 13000199, 13000200, 13000201, 13000202, 13000203, 13000204, 13000205, SN13000206, 14000004, 14000005, 14000006, 14000008, 14000009, 14000010, 14000011, 14000013, 14000014, 14000015, 14000016, 14000017, 14000019, 14000022, 14000024, 14000026, 14000028, 14000029, 14000031, 14000032, 14000033, 14000034, 14000037, 14000038, 14000041, 14000043, 14000046, 14000047, 14000048, 14000049, 14000050, 14000051, 14000052, SN14000061, 14000069, SN14000072, 14000075, 14000077, 14000078, 14000079, 14000080, 14000081, 14000082, 14000084, 14000086, 14000087, 14000088, 14000089, SN14000096, SN14000097, SN14000098, SN14000099, SN14000100, 14000092, 14000083, 14000085, 14000103, 14000106, 14000083, 14000109, 14000112, 14000116, 14000119, 14000118, 14000120, 14000121, 14000122, 14000124, SN14000128, 14000127, 14000126, SN14000130, 14000134, 14000135, 14000138, 14000139, SN15000002, 14100044, 14100045, 14100053, 14100054, 14100085, 14100086, 14100087, 14100088, 14100090, 14100094, 14100095, 14100097, 14100101, 14100104, 14100105, 14100109, 14100111, 14100117, 14100126, 14100127, 14100128, SN14100046, 15100019, 15100023, SN14100047, SN14100055, SN14100071, SN14100072, SN14100073, SN14100074, SN14100075, 15100028, 15100034, 15100035, 15100036, 15100039, 15100040, 15100041, 15100042, 15100047, 15100048, 15100050, 15100071, 15100084, 15100087, 15100088, 15100089, 15100091, 15100092, 15100093, 15100094, 15100095, 15100097, 15100098, 15100101, 15100103, 15100104, SN15100020, SN15100009, SN15100016, SN15100010, SN15100011, SN15100013 and SN15100018.

Distribution pattern

Nationwide Distribution including AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NM, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI.