device · product 1 of 1
TOPAZ MicroDebrider 0.8 mm REF AC4040-01; TOPAZ MicroDebrider IFS 0.8 mm REF ACH4040-01; TOPAZ MicroDebrider 0.8 mm REF Q6000-01; TOPAZ EPF MicroDebrider 0.8 mm 45 REF Q6002-01
- Recall number
- Z-1321-2016
- Initiated
- March 03, 2016
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- ArthroCare Corporation
- Quantity
- 26,903 devices
App-derived interpretation
During functional testing of devices after real time aging, it was determined that product failed to meet shelf-life at the expiration date indicated on the product labels.
Official device-enrichment evidence · Sourced
Package design/selection
Inspect official wording and provenance
Reason for recall
During functional testing of devices after real time aging, it was determined that product failed to meet shelf-life at the expiration date indicated on the product labels.
Code information
1031506, 1031507, 1032123, 1032325, 1032741, 1032742, 1033443, 1033965, 1033966, 1034397, 1039120, 1041765, 1044089, 1044090, 1045862, 1048237, 1048933, 1049185, 1052071, 1056212, 1060586, 1060979, 1063394, 1064599, 1067696, 1067748, 1069174, 1071358, 1074125, 1075714, 1070521, 1070812, 1071974, 1078025, 1077499, 1082811, 1084095, 1084094, 1085458, 1086436, 1086437, 1087917, 1094313, 1097573, 1103824, 1107183, 1107297, 1113562, 1113563, 1114534, 1114535, 1114536, 1115896, 1031518, 1032308, 1032309, 1032310, 1032755, 1032756, 1033972, 1033973, 1034225, 1034226, 1035283, 1035284, 1038547, 1043507, 1046556, 1047064, 1048722, 1049266, 1051812, 1052070, 1054443, 1055730, 1059865, 1063661, 1067691, 1067273, 1067747, 1068019, 1069172, 1071357, 1072853, 1072854, 1070497, 1070808, 1071971, 1076849, 1077450, 1079140, 1079220, 1078115, 1081753, 1082361, 1082810, 1086380, 1094312, 1094312, 1094721, 1097571, 1101041, 1107295, 1109045, 1111935, 1111936, 1113534, 1113535, 1113536, 1114532, 1115888, 1115891, 1116959, 1116130, 1116720, 1117209, 1117214, 1031341, 1032311, 1033761, 1034446, 1035691, 1035692, 1041775, 1043962, 1046557, 1048236, 1050067, 1052651, 1057008, 1058808, 1059867, 1063665, 1064594, 1067695, 1068021, 1071319, 1071320, 1073441, 1074876, 1077498, 1080520, 1114533, 1115893, 1031340, 1032962, 1033266, 1045359, 1047065, 1051331, 1056849, 1064592, 1067694, 1070709
Distribution pattern
Worldwide Distribution-US (nationwide) and the following countries-Austria, Belgium, Bulgaria, Canada, Denmark, Estonia, Finland, France, Germany, Gibraltar, Great Britain, Greece, Ireland, Israel, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, China, Hong Kong, India, Lebanon, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arabic Emirates, Vietnam, Australia, Japan, New Zealand, Brazil, and Chile.