Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 73761

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 19, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Elekta, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Radionuclide Radiation Therapy System Product The product is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters.

Z-1561-2016
Recall number
Z-1561-2016
Initiated
April 19, 2016
Classification
Class II
Status
Terminated
Recalling firm
Elekta, Inc.
Quantity
217

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The latches of the frame adapter can be locked even if the locating pins of the frame adapter is not inserted into the corresponding holes in the coordinate frame.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The latches of the frame adapter can be locked even if the locating pins of the frame adapter is not inserted into the corresponding holes in the coordinate frame.

Code information

Perfexion - 14753-099/6052, 12579-007/6250, 10394-LGK01/6144, 11920-C5210/6088, 17397-001/6158, 105528-001/6122, 14857-001/6028, 12236-006/6097, 12473-003/6005, 14857-001/6028, 16867-001/6166, 12668-003/6145 10208-003/6215, 10698-001/6237, 10979-002/6213, 10316-500/6210, 21061-001/6195, 20364-001/6103, 11720-002/6143, 12620-002/6120, 10611-001/6081, 12846-004/6064, 11524-002/8001, 11420-001/6212, 11420-001/6228, 14983-001/6076, 11083-HOM001/6070, 10247-002/6136,10684-002/6090,11868-004/6002, 10795-K08002/6063, 20354-ENS/6160, 11792-005/6077, 11291-001/6018, 10326-003/6105, 18630-001/6031, 17322-001/6121, 30000656-002/6227, 12629-003/6026, 10924-001/6089, 11154-003/6050, 20438-003/6013 17522-001/6214, 17525-002/6038, 11414-LGK002/6205, 11940-LGK002/6181, 10514-LGK002/6183, 11639-LGK001/6042, 11390-LGK001/6109, 10416-LGK001/6125, 11670-LGK001/6133, 11990-LGK001/6118, 10648-LGK001/6098, 15104-LGK001/6080, 10654-LGK001/6096, 10959-LGK001/6132, 10997-LGK001/6087, 15748-LGK001/6106, 11149-LGK002/6206, 11986-LGK002/6218, 10929-LGK002/6175, 10879-LGK002/6234, 11434-LGK002/6255, 11991-LGK002/6198, 13480-LGK002/6148, 10425-LGK002/6254, 10583-LGK002/6161, 11563-LGK002/6162, 10103-LGK02/6149, 10541-LGK002/6142, 10848-LGK001/6108, 11433-LGK001/6116, 14900-LGK001/6058, 15045-LGK001/6091 11784-LGK001/6055, 30000732-001/6223, 40017-002/6170, 12085-006/6163, 12109-002/6169, 12959-002/6033, 10209-001/6046, 10960-005/6193, 0002718-001/6113, 30001751-WOJ701/6068, 12451-001/6019, 14878-003/6067, 105618-001/6110, 19634-001/6155, 16647-001/6187, 14891-001/6104, 30001527-001/6231, 12028-LGK002/6217, 11586-001/6157, 10728-001/6062, 11999-001/6126, 15416-001/6086, 10459-001/6123, 10158-001/6177, 10234-002/6111, 40023-001/6178, 10444-001/6085, 17033-003/6043, 11530-001/6101, 11553-001/6024, 10779-07/6069, 13465-CHUV03/6094, 10424-001/6182, 13370-002/6075, 12214-003/6007, 30000655-001/6199, 10152-001/6159, 11550-001/6065, 10118-001/6230, 12766-001/6112, 13160-001/6039, 11042-001/6003, 12270-001/6017, 18062-001/6146, 11710-004/6141, 11820-006/8006, 11570-004/6220, 11127-009/6219, 12164-004/6221, 10434-004/6244, 10542-012/6238, 13826-002/6239, 10567-041/6191, 98300-003/6140, 10567-037/6119, 14569-001/6135, 10643-005/6117, 11798-009/6060, 17357-001/6130, 11813-003/6034, 10356-007/6079, 10567-045/6232, 10567-032/6074, 10567-030/6036, 10567-025/6029, 11293-004/6004, 12269-002/6082, 11772-004/6048, 10164-004/6009, 10349-003/6066, 10567-026/6011, 10560-003/6127, 11073-008/6020, 10567-036/6102, 14236-001/6114, 10567-035/6092, 10379-002/6061, 11030-006/6030, 13050-001/6037, 10909-009/6010, 11714-003/6072, 12985-003/6025, 14030-002/6015, 10567-031/6054, 10064-008/6021, 13558-001/6099, 10567-038/6107, 11939-003/6008, 10591-002/6083, 10567-024/6014, 11848-003/6095, 11624-003/6078, 10383-007/6100, 11395-010/6032, 11198-005/6124, 10982-003/6053, 11256-001/6012, 10905-004/6016, 10387-011/6093, 11138-004/6115, 10524-003/6073, 10908-003/6045, 10762-005/6057, 10193-007/6056, 10673-016/6006, 11007-005/6035, 11791-006/6022, 11377-003/6186, 12752-004/6242, 11039-006/6188, 10597-002/6167, 17539-001/6185, 14153-001/6197, 30000118-001/6240, 10567-039/6180, 15053-001/6179, 11731-005/6200, 11090-005/6165, 11077-007/6207, 30000171-002/6172, 10760-006/6173, 11733-001/6192, 10817-009/6204, 13514-001/6211, 10877-GK01/6151, 10436-009/6150, 11396-003/6176, 10917-004/6128, 11373-002/6184, 10567-042/6129, 10567-043/6203, 12614-003/6139, 10297-006/6171, 11347-003/6134 LEKSELL GAMMA KNIFE ICON - 11868-008/6131, 11290-ENS1/6174, 12027-EZT-G1/6041, 10444-LGK004/8003, 11352-001/6044, 10622-005/6194, 10886-013/6027, 11201-004/6047, 11834-003/6222

Distribution pattern

Worldwide Distribution - US Nationwide in the states of AK, AZ, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MI, MN, MS, MO, MT, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, WA, WV, WI, and countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Italy, Japan, Jordan, Malaysia, Mexico, Morocco, Netherlands, Norway, Philippines, Poland, Portugal, Russia, Saudi Arabia, Serbia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom