Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 73815

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 21, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Angiodynamics Inc. (Navilyst Medical Inc.)

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Vortex MEDICAL AngioVac Circuit, REF/UPN VTX3100, STERILE, Rx Only, Single Use Only -- The AngioVac Circuit and AngioVac Circuit Pack are designed to be used with the AngioVac Cannula.

Z-1701-2016
Recall number
Z-1701-2016
Initiated
March 21, 2016
Classification
Class II
Status
Terminated
Quantity
Domestic: 339 kits

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Maquet Cardiopulmonary, the manufacturer of the RotaFlow Centrifugal Pump, determined that the products may possess levels of endotoxin that exceed regulatory limits. AngioDynamics confirmed that the affected RotaFlow Centrifugal Pumps were included in packaged AngioVac Circuit Packs.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Environmental control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Maquet Cardiopulmonary, the manufacturer of the RotaFlow Centrifugal Pump, determined that the products may possess levels of endotoxin that exceed regulatory limits. AngioDynamics confirmed that the affected RotaFlow Centrifugal Pumps were included in packaged AngioVac Circuit Packs.

Code information

Lot Numbers 37FE2901, 37GD2908, 37KD2405

Distribution pattern

Nationwide Distribution.

device · product 2 of 2

AngioVac Circuit Pack (Tandem Packs), REF/Catalog No. 25-187, UPN H965251870, STERILE, Rx Only, Single Use Only -- The AngioVac Circuit and AngioVac Circuit Pack are designed to be used with the AngioVac Cannula.

Z-1702-2016
Recall number
Z-1702-2016
Initiated
March 21, 2016
Classification
Class II
Status
Terminated
Quantity
Domestic: 377 kits

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Maquet Cardiopulmonary, the manufacturer of the RotaFlow Centrifugal Pump, determined that the products may possess levels of endotoxin that exceed regulatory limits. AngioDynamics confirmed that the affected RotaFlow Centrifugal Pumps were included in packaged AngioVac Circuit Packs.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Environmental control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Maquet Cardiopulmonary, the manufacturer of the RotaFlow Centrifugal Pump, determined that the products may possess levels of endotoxin that exceed regulatory limits. AngioDynamics confirmed that the affected RotaFlow Centrifugal Pumps were included in packaged AngioVac Circuit Packs.

Code information

Lot Numbers: 4876386, 4889699, 4889700, 4898652, 4902111, 4902117, 4906576, 4941511, 4941511D, 4943459, 4955686, 4956929, 4969475, 4975493, 4991710, 4999714.

Distribution pattern

Nationwide Distribution.