device · product 1 of 1
SOMATOM Force, Computed Tomography x-ray system intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission.
- Recall number
- Z-1558-2016
- Initiated
- April 01, 2016
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 37 CTsystems
App-derived interpretation
The Neonate Head protocol with the pediatric kernel Hp38 could result in artefacts and possibly lead to a misdiagnosis (i.e. either non-existing blood or liquid is mimicked in the images or actual existing blood or liquid is not depicted as expected). There is also a risk of a potential misdiagnosis in using this protocol for surgery planning. Investigations revealed an incorrect parameterization of the reconstruction algorithm applied for the pediatric head kernel as cause of the problem.
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
The Neonate Head protocol with the pediatric kernel Hp38 could result in artefacts and possibly lead to a misdiagnosis (i.e. either non-existing blood or liquid is mimicked in the images or actual existing blood or liquid is not depicted as expected). There is also a risk of a potential misdiagnosis in using this protocol for surgery planning. Investigations revealed an incorrect parameterization of the reconstruction algorithm applied for the pediatric head kernel as cause of the problem.
Code information
SOMATOM Force Material # 10742326, serial numbers 75550, 75526, 75512, 75567, 75536, 75514, 75563, 75510, 75454, 75540, 75571, 75437, 75500, 75527, 75475, 75535, 75481, 75439, 75450, 75482, 75460, 75493, 75476, 75458, 75528, 75467, 75487, 75478, 75513, 75532, 75555, 75524, 75515, 75559, 75501, 75570, 75576, 75450, 75482, 75528, 75576
Distribution pattern
US Nationwide Distribution : NC, NH, IL, MA, WA, PA, MD, MN, TX, NY, SC, OH, CA, KY, MT, IA