Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 73884

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 06, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Nuvectra

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Nuvectra, Algovita Spinal Cord Stimulation System Clinical Programmer, Model 4500. For use with Algovita Spinal Cord Stimulation Systems. The version of the software in the Clinician Programmer is v1.1.5. The Algovita Spinal Cord Stimulation system consists of a stimulator (EPG or IPG) that is physically and electrically connected to one or more leads inserted into the patient¿¿"s spinal epidural space. The lead delivers electrical stimulation originating at the stimulator with the purpose of blocking pain signals going to the patient¿¿"s brain. The stimulators are rechargeable, and the frequency of recharging is dependent on individual patient¿¿"s use of the system. The Clinician Programmer (CP) is a hand-held, touch screen device used to create and adjust stimulation parameters that are developed to optimize the therapy for each patient. The CP communicates with the stimulator via the Medical Implant Communication Service (MICS). The CP wirelessly transfers the stimulation parameters to the stimulator. In most clinical settings, the CP can be used by the Health Care Professional to program multiple patient EPGs and IPGs

Z-1600-2016
Recall number
Z-1600-2016
Initiated
April 06, 2016
Classification
Class II
Status
Terminated
Recalling firm
Nuvectra
Quantity
17 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
If used, the optional Swap feature copies programming parameters from the external pulse generator (EPG) as well as unintended calibration data to the implantable pulse generator (lPG). This may result in the user being unable to recharge the lPG. This safety notice only applies to CPs Model 4500.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

If used, the optional Swap feature copies programming parameters from the external pulse generator (EPG) as well as unintended calibration data to the implantable pulse generator (lPG). This may result in the user being unable to recharge the lPG. This safety notice only applies to CPs Model 4500.

Code information

Serial 000010 000011 FF0043 FF0045 000014 000015 000016 FF0054 FF0048 FF004B FF005D FF0064 FF0065 FF0067 FF00A9 FF00AD FF00BB Lot P178713 P185700 P187989 P195875 P213108 P257056

Distribution pattern

Germany