VITROS 3600 Immunodiagnostic System, Catalog Number 6802783,for use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents.
- Recall number
- Z-1913-2016
- Initiated
- April 06, 2016
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics
- Quantity
- US: 218 units, Foreign: 587 units
App-derived interpretation
software timing anomaly
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
Ortho Clinical Diagnostics confirmed a software timing anomaly in which the VITROS 3600 and 5600 Systems may unexpectedly aspirate from or dispense into an unintended sample that may lead to erroneous or contaminated results.
Code information
Software Version 3.2.2 & below; Serial Numbers J36000012 - J36000933.Unique Device Identifier No. 10758750002979, IVD.
Distribution pattern
Nationwide, Puerto Rico and Foreign distribution to the following countries: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.