device · product 1 of 1
Syngo Dynamics; Kinetdx Picture Archiving and Communications System
- Recall number
- Z-1601-2016
- Initiated
- April 14, 2016
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 84 units
App-derived interpretation
Siemens has identified changes in default carry forward workflow using VA10B_HF03 version of Syngo Dynamics. If the template design has measurements inside a carry forward-configured report section, this will result in carrying forward the entire section of the report including data from previous studies. This may result in the physician making analysis based on the old measurements.
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
Siemens has identified changes in default carry forward workflow using VA10B_HF03 version of Syngo Dynamics. If the template design has measurements inside a carry forward-configured report section, this will result in carrying forward the entire section of the report including data from previous studies. This may result in the physician making analysis based on the old measurements.
Code information
Model Numbers: 10091637, 10091673, 10091805, 10091807 Serial Numbers: 85021, 85040, 85042, 85045, 85047, 85057, 85079, 85080, 85081, 85086, 85090, 85115, 85143, 85145, 85146, 85147, 85149, 85157, 85167, 85168, 85170, 85172, 85177, 85178, 85185, 85186, 85189, 85191, 85234, 85235, 85239, 85281, 85307, 85309, 85322, 85325, 85326, 85329, 85335, 85338, 85350, 85353, 85354, 85368, 85370, 85374, 85377, 85380, 85385, 85387, 85401, 85418, 85421, 85422, 85427, 85435, 85442, 85447, 85462, 85467, 85468, 85469, 85470, 85473, 85481, 85482, 85484, 85859, 85860, 85866, 85867, 85871, 85872, 85873, 85874, 85875, 85876, 85880, 85883, 85884, 85893, 85896, 85910, 85918
Distribution pattern
Nationwide: AZ, CA, IL, IN, MA, MD, MI, MN, NJ, NY, OH, OR, PA, TN, TX, UT, WA, WI