Event 73974

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 20, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
GE Medical Systems, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

GE, Revolution CT. The system is intended for head, whole body, cardiac, and vascular X-ray Computed Tomography applications

Z-1589-2016
Recall number
Z-1589-2016
Initiated
April 20, 2016
Classification
Class II
Status
Terminated
Recalling firm
GE Medical Systems, LLC
Quantity
26 (US = 4;OUS = 22)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
An out of tolerance torque tool was used on three joints during the manufacturing process. If the X-ray Tube becomes loose on the CT gantry during operation, this could result in beam tracking errors, potential scan aborts and/or a possible image artifact. There is a secondary safety mechanism to prevent components from being expelled from the gantry if the bolts fail.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An out of tolerance torque tool was used on three joints during the manufacturing process. If the X-ray Tube becomes loose on the CT gantry during operation, this could result in beam tracking errors, potential scan aborts and/or a possible image artifact. There is a secondary safety mechanism to prevent components from being expelled from the gantry if the bolts fail.

Code information

Mfg Lot or Serial # System ID 442706CN8 727398REVO 443227CN4 940898REVO 00000442994CN0 919684REVCT 443069cn0 404616CTR 00000442968CN4 83141966438 443342CN1 905521RCT1 00000444648CN0 82421100836 00000442479CN2 82421190324 00000443071CN6 82421300072 00000443308CN2 82421040370 00000444206CN7 82421160425 00000442792CN8 M5480791 00000443675CN4 410014CT04 00000443869CN3 A51773126 00000442785CN2 REV015 00000443501CN2 REV013 00000442940CN3 850210884 00000443590CN5 850210890 00000442289CN5 NO1031CT01 00000444156CN4 CT444156CN4 00000442748CN0 CT442748CN0 00000443952CN7 R002CT09 00000443540CN0 A201CT03 00000444038CN4 AE1015CT02 00000443409CN8 00145CTS01 00000443994CN9 00095CTS04

Distribution pattern

Worldwide Distribution - US including FL, GA, NC, and TX and Internationally to AUSTRALIA, CANADA, CHINA, FRANCE, GERMANY, ITALY, JAPAN, REPUBLIC OF KOREA, NORWAY, PERU, SWEDEN UNITED ARAB EMIRATES, UNITED KINGDOM.