ADVIA Centaur XPT Immunoassay System
- Recall number
- Z-0072-2017
- Initiated
- April 28, 2016
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 292 units Total (3 domestically & 289 internationally)
App-derived interpretation
The ADVIA Centaur¿ XPT default setting for the Daily Maintenance Task (Daily Cleaning Procedure) frequency may have the Daily Maintenance Task frequency set to, As needed instead of Daily. Not performing the daily maintenance may impact any assay and can be detected through monitoring of quality control and calibration results. Systems running in a language other than English are not impacted.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
The ADVIA Centaur¿ XPT default setting for the Daily Maintenance Task (Daily Cleaning Procedure) frequency may have the Daily Maintenance Task frequency set to, As needed instead of Daily. Not performing the daily maintenance may impact any assay and can be detected through monitoring of quality control and calibration results. Systems running in a language other than English are not impacted.
Code information
None
Distribution pattern
Worldwide Distribution - US including MD, MN, & MS; **Internationally** AE, AT, AU, BE, BR, CA, CN, DE, DK, EG, ES, FI, FR, GB, GP, HU, IN, IT, JO, JP, KR, MY, NO, NZ, PH, PT, QA, SA, SE, SG, TR & VN