Event 74567

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 05, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Beckman Coulter Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Allegra X-5 rotor yoke, Catalog No. (Instruments) B30589 and B30590 (Rotor) B30593. Product Labeling: IFU B29071AA. Product Usage: For In Vitro Diagnostic Use. The Allegra X-5 is intended for the separation of components through the use of relative centrifugal force. It is designed to separate human samples, including blood and other body fluids, for processing, analysis and in vitro diagnostic testing, as well as non-human body samples and chemicals, including industrial and environmental samples. This centrifuge should be operated by qualified personnel only.

Z-2274-2016
Recall number
Z-2274-2016
Initiated
February 05, 2016
Classification
Class II
Status
Terminated
Recalling firm
Beckman Coulter Inc.
Quantity
62 units total (33 units in US)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A defective rotor yoke on Allegra X-5 allows improper rotor seating on the hub resulting in possible rotor failure if the buckets are loaded in an imbalanced configuration.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A defective rotor yoke on Allegra X-5 allows improper rotor seating on the hub resulting in possible rotor failure if the buckets are loaded in an imbalanced configuration.

Code information

AFB13H005, AFB13H008, AFB13H012, AFB13K002, AFB13K005, AFB13K008, AFB13J002, AFB13J003, AFB13J004, AFB13J005, AFB13J007, AFB15J001, AFB15C010, AFB13K009, AFB14H001, AFB15C008, AFB15C003, AFB15C005, AFB15C004, AFB15C006, AFB14H005, AFB14H003, AFB13K004, AFB13K006, AFB13K007, AFB13J008, AFB13K001, AFB13J006, AFB13J010, AFB13H013, AFB13H009, AFB13H003, AFB13H004

Distribution pattern

Worldwide Distribution - US Nationwide (including Puerto Rico) and the countries of: Australia, Canada, Colombia, Germany, Guatemala, Netherlands, Singapore, Tunisia, , Spain