Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 74593

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 22, 2016
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Olympus Scientific Solutions Americas

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Olympus Scientific Solutions Americas Corporation (OSSA) Model: Analytical X-Ray System X -Ray Fluorescence. Intended to emit ionizing radiation Delta with models: DC-2000, DC-4000, DC-6000, DI-2000, DP-2000, DP-4000, DP-4050, DP- 6000, DPO-4000, DPO-6000, DS-2000, DS-4000, DS-6000, DS-6500, Delta, Delta 50, Delta Classic Plus, Delta Element, Delta Premium , Delta Professional , and Delta-Pl us.

Z-2255-2016
Recall number
Z-2255-2016
Initiated
June 22, 2016
Classification
Class II
Status
Ongoing
Quantity
2, 210 units distributed in the USA

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Trigger lock option does not always prevent x-ray generation. Specifically if the unit is in test mode or repeat test mode then the trigger lock may not prevent x-ray initiation when the trigger is pulled. The correction is via repair by a software update.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Radiation Control for Health and Safety Act

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Trigger lock option does not always prevent x-ray generation. Specifically if the unit is in test mode or repeat test mode then the trigger lock may not prevent x-ray initiation when the trigger is pulled. The correction is via repair by a software update.

Code information

All Models of the Delta XRF Analyzer with Software version 2.5.17.1 and higher

Distribution pattern

Nationwide Distribution.