device · product 1 of 1
Syngo.via, picture archiving and communications system software controlled. Intended to be used for viewing, manipulation, communication, and storage of medical images.
- Recall number
- Z-2245-2016
- Initiated
- June 23, 2016
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 50 units
App-derived interpretation
Incorrect values for the volume calculation. Software
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
Incorrect values for the volume calculation. Software update VB30B via Update Instructions SY018/16/P to resolve software errors.
Code information
310289,130542,130322,310149,130325,221063,310168,101762,221048,221096,220783,310195,130451,400031,130292,310114,220846,130269,130163,130402,130826,221051,102336,310282,220888,10007,221052,221049,130518,221002,221091,220915,130520,130472,100469,310249,130122,100013,130378,310188,102453,310079,310080,130240,130382,220985,310171,220879,220995, and 310268.
Distribution pattern
Nationwide Distribution -- AR, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TX, and UT.