Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 74634

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 23, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Stryker Howmedica Osteonics Corp.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Stryker Orthopaedics Modular Handle Triathlon Instruments, Non-Sterile, REF 6541-4-808, Lot 123, NDC M546654148080S Product Usage: Is a reusable instrument that is part of the Triathlon Primary total knee arthroplasty (TKA) system as well as the Triathlon Partial Knee Replacement (PKR) system. The use of the handles in Primary TKA is optional and may be used to: (1) Assist in stabilizing the Triathlon Femoral Sizer, (2) Aid in stabilization during use and removal of the Triathlon 4:1 Cutting Block and (3) Aid in assembly of the Anterior MIS Distal Resection Guide to the Anterior Skim Cut Guide. Per the Triathlon PKR surgical protocol, the handle is required for a PKR surgery and is used to: (1) Assemble and disassemble the Femoral Trial Drill Guide into the Femoral Trial and (2) Remove the Peg Trial from the Femoral Trial.

Z-2377-2016
Recall number
Z-2377-2016
Initiated
June 23, 2016
Classification
Class II
Status
Terminated
Quantity
12,469 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Stryker Orthopaedic is voluntarily recalling the Triathlon Modular Handle because analysis revealed that the press fit specifications between the dowel pin and the mating hole in the shaft were not being achieved potentially resulting in disassociation from the instrument.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Stryker Orthopaedic is voluntarily recalling the Triathlon Modular Handle because analysis revealed that the press fit specifications between the dowel pin and the mating hole in the shaft were not being achieved potentially resulting in disassociation from the instrument.

Code information

Catalog No. 6541-4-808 Lot Nos.: SB1K14, SB1K14A, SB1K14A1, SB1K14P, SB1K15, SB1K15A, SB1K15A1, SB1K15A2, SB1M25, SB1M25A, SB1M25A1, SB1M25A2, SB1M25A2X1, SB1N02, SB1N02A, SB1N02A1, SB1N02A2, SB1N02A3, SB1N02X1, SB1T05, SB1T05A, SB1T05A1, SB1T05AX1, SB1T05AX2, SB1T05KSB1V01, SB1V01K, SB1V01L, SB1V01X1, SB1V01X2, SB1V01X3, SB1W14, SB1W14K, SB1W14X1, SB2T19, SB2T19A, SB2T19D, SB2T19DX1, SB2T19DX2, SB2T19J, SB2T19K, SB2T19KX1, SB2T19KX2, SBT19X3, SB2V87, SB2V87A, SB2V87J, SB2V87JX1, SB2V87K, SB2V87X, SB2V87X1, SB2W23, SB2W23G, SB2W23X, SB2W23X1, SB3A20, SB3A20A, SB3A21, SB3A21D, SB3A21M, SB3A21X, SB3A22, SB3A22A1, SB3A22X1, SB3A55, SB3A55A, SB3A55A2, SB3A55AX1, SB3A55M, SB3A55X1, SB3H25, SB3H25D, SB3H25G, SB3H25X1, SB3H26, SB3H26D, SB3H26DX1, SB3H26X2, SB3K09, SB3K09X1, SB3K10, SB3K10L, SB3K10M, SB3K10MX1, SB3K10X1, SB3L19, SB3L19D, SB3L19K, SB3L19L, SB3L19P, SB3L19P1, SB3L20, SB3L20K, SB3L45, SB3L45A, SB3L45A1, SB3L45A2, SB3L45D, SB3L45DX1, SB3L45L, SB3L45T, SB3M17, SB3M17A, SB3M17K, SB3M17M, SB3M18, SB3M18A, SB3M18AX1, SB3M18D, SB3M18M, SB3M18R, SB3M19, SB3M19A, SB3M19A1, SB3M19S, SB3M19T, SB3M19W, SB3N46, SB3N46D, SB3N46P, SB3N46X1, SB4C12, SB4C12X1, SB4C37, SB4C37X1, SB4C38, SB4C38A, SB4C38D, SB4E18, SB4E181, SB4E181X1, SB4E18A, SB4E18D, SB4E18T, SB4H79, SB4H79A, SB4H79M, SB4L05, SB4L05A, SB4N17, SB4N17E, SB4N17M, SB4N17M1, SB4N18, SB4N18A, SB4N18A1, SB4N18D, SB4N18M, SB4N18T, SB4S01, SB4S01D, SB4S01J, SB4V07, SB4V07L, SB4V07M, SB4V07MM, SB5A02, SB5A02A, SB5A02A1, SB5A55

Distribution pattern

Worldwide Distribution - US Nationwide