device · product 1 of 1
Carestream Touch Prime, Catalog Number 1738830, and Carestream Touch Prime XE, Catalog Number 1738822 --- Common Name: Touch Ultrasound Diagnostic ultrasound imaging or fluid flow analysis of the human body
A software issue related to the generic volume measurement functionality could result in the volume of the subject anatomy being overestimated. For example, measurement results of the anatomy may appear enlarged or distended when it actually is not.
Code information
Touch Prime Serial Numbers: 1037, 1038; -- Touch Prime XE Serial Numbers: 1021, 1079, 1067, 1041, 1043
Distribution pattern
Worldwide Distribution: US Distribution to states of: GA, IA, and TX; and country of: Italy.