device · product 1 of 1
The Connecting Bracket (Fixture) Device (1002.65A0) is designed as a connecting element to support ancillary head rests on a Maquet operating room table immediately before, during and after surgical interventions as well as for examination and treatment. The fixture is used to adapt a range of different head and horseshoe head rests with a central square mount
Post Market Surveillance cases reported to Maquet showing instances in which the welded joint between the square mount and the first fixable swivel joint of Maquet's Connecting Bracket (Fixture) Device is broken.
These labels are deterministic app interpretations, not FDA categories.
Post Market Surveillance cases reported to Maquet showing instances in which the welded joint between the square mount and the first fixable swivel joint of Maquet's Connecting Bracket (Fixture) Device is broken.
Code information
Part Number/Catalog Number 1002.64A0
Distribution pattern
Nationwide Distribution to CA, FL, IA, IN, MO, UT and WI