Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 75124

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 01, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Medtronic Navigation, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Visualase Visualization Stylets PN 020-2301. The individual components within the Visualase Body Accessory Kit are packaged in Tyvek pouches. The packaged devices are then placed into a cardboard box which is labeled with the Body Accessory Kit label. Labeled sterile. The visualization stylet is constructed of titanium and used during Visualase laser ablation procedures by inserting it into intracorporeal catheter lumens for improved catheter stiffness and MRI visualization during procedures

Z-0322-2017
Recall number
Z-0322-2017
Initiated
September 01, 2016
Classification
Class II
Status
Terminated
Quantity
100 total units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Mixed-up of materials/components

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field.

Code information

Lot number 151481

Distribution pattern

Worldwide Distribution - US including CA, GA, FL, and TX. Internationally to Australia.

device · product 2 of 2

Visualase Body Accessory Kit PN 020-3000. The individual components within the Visualase Body Accessory Kit are packaged in Tyvek pouches. The packaged devices are then placed into a cardboard box which is labeled with the Body Accessory Kit label. Labeled sterile. The visualization stylet is constructed of titanium and used during Visualase laser ablation procedures by inserting it into intracorporeal catheter lumens for improved catheter stiffness and MRI visualization during procedures.

Z-0323-2017
Recall number
Z-0323-2017
Initiated
September 01, 2016
Classification
Class II
Status
Terminated
Quantity
100 total units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Mixed-up of materials/components

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field.

Code information

Lot number 151591

Distribution pattern

Worldwide Distribution - US including CA, GA, FL, and TX. Internationally to Australia.