Event 75157

4 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 24, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Medical Systems

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

4 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 4

Brilliance 64 CT Model Number 728231 Product Usage: Computed Tomography X-ray systems intended to product cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Z-0009-2017
Recall number
Z-0009-2017
Initiated
August 24, 2016
Classification
Class II
Status
Terminated
Recalling firm
Philips Medical Systems
Quantity
185

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults to standard resolution which doesn't experience the issue.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults to standard resolution which doesn't experience the issue.

Code information

4001, 4016, 5169, 9056, 9071, 9074, 9099, 9112, 9134, 9251, 9280, 9367, 9515, 9529, 9538, 9541, 9548, 9549, 9554, 9559, 9561, 9573, 9574, 9600, 9614, 9618, 9625, 9631, 9638, 9640, 9646, 9653, 9565, 9659, 9667, 9670, 9678, 9686, 9688, 9689, 9703, 9707, 9711, 9720, 9738, 9740, 9755, 9763, 9781, 9782, 9802, 9805, 9811, 9822, 9833, 9837, 9838, 9859, 9859, 9862, 9864, 9876, 9890, 9898, 9901, 9909, 9910, 9917, 9920, 9922, 9923, 9927, 9930, 9933, 9983, 9989, 9993, 9996, 9997, 9910, 9917, 9920, 9922, 9923, 9927, 9930, 9933, 9983, 9989, 9993, 9993, 9997, 10012, 10014, 10021, 10024, 10032, 10042, 10047, 10050, 10051, 10077, 10092, 10099, 10106, 10150, 10179, 10180, 10193, 10206, 10207, 10208, 10213, 10216, 10221, 10250, 10255, 10299, 10302, 10379, 10394, 10422, 10446, 10470, 10473, 10503, 10514, 10605, 10610, 10671, 10698, 10743, 29002, 29005, 29014, 29015, 29049, 29056, 29057, 29107, 29115, 90098, 90106, 90124, 90131, 90154, 90205, 91003, 95022, 95024, 95130, 95157, 95191, 95197, 95213, 95222, 95236, 95239, 95244, 95258, 95261, 95341, 95343, 95345, 95351, 95364, 95374, 95387, 95399, 95409, 95410, 94539, 95441, 95475, 95496, 95504, 95507, 95512, 95524, 95538, 95551, 95563, 95614, 95616, 95656, 95668, 95688, 123029, 95691, 95707, 95811, 122921, 122921, 123092, 123096, 123116, 123117, 6003368, 95341A

Distribution pattern

Worldwide Distribution - US Nationwide in the states of:: AK, AL, AR, AZ, CA, CO, CT, DE, GA, IA, IL, IN, FL, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NJ, NH, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, & WY. and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Brazil, Cambodia, China, COlumbia, Cuba, Denmark, Dijbouti, Dominican REpublic, Egypt, Estonia, Finland, France, Germany, Gilbraltar, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Libya, Lithuania, Malaysia, Martinique, Monaco, Mongolia, Mozambique, Myanmar, Netherlands, New Zealand, Norway, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russian Federation, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Seitzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uniter Arab Emirates, United Kingdom, Uzbekistan, and Vietnam.

device · product 2 of 4

Ingenuity Core Model Number 728321 Product Usage: Computed Tomography X-ray systems intended to product cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Z-0010-2017
Recall number
Z-0010-2017
Initiated
August 24, 2016
Classification
Class II
Status
Terminated
Recalling firm
Philips Medical Systems
Quantity
358

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults to standard resolution which doesn't experience the issue.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults to standard resolution which doesn't experience the issue.

Code information

30041, 31001, 33084, 52000, 52004, 52007, 52008, 52010, 52012, 52013, 52014, 52015, 52018, 52019, 52020, 52021, 52022, 52024, 52025, 52026, 52027, 52029, 52030, 52031, 52032, 52033, 52034, 52035, 52036, 52037, 52038, 52039, 52040, 52041, 52042, 52043, 52044, 52045, 52046, 52047, 52048, 52050, 25052, 52053, 52054, 52055, 52056, 52057, 52058, 52059, 52061, 52063, 52064, 52065, 52067, 52068, 52069, 52070, 52071, 52072, 52073, 52074, 52075, 52076, 52077, 52078, 52079, 52080, 52081, 52082, 52083, 52084, 132164, 132465, 310001, 310002, 310003, 310009, 310017, 310019, 310022, 310023, 310026, 310032, 310039, 310042, 310052, 310054, 310055, 310059, 310064, 310069, 310070, 310071, 310081, 310087, 310088, 310090, 310096, 310098, 310105, 310107, 310111, 310113, 310116, 310119, 310121, 310125, 310129, 310133, 310134, 310135, 310141, 310143, 310149, 310150, 310151, 310157, 310159, 310160, 310170, 310203, 310204, 310205, 310206, 310207, 310208, 310209, 310210, 310211, 310212, 310213, 310214, 310215, 310216, 310217, 310218, 310219, 310220, 310221, 310222, 310223, 310224, 310225, 310226, 310227, 310228, 310229, 310230, 310231, 310232, 310233, 310234, 310235, 310236, 310237, 310238, 310239, 310240, 310241, 310242, 310243, 310244, 310245, 310246, 310247, 310248, 310249, 310250, 310251, 310252, 310253, 310254, 310255, 310256, 310257, 310258, 310259, 310260, 310261, 310262, 310263, 310264, 310265, 310266, 310267, 310268, 310269, 310270, 310271, 310272, 310273, 310274, 310275, 310276, 310277, 310278, 310279, 310280, 310281, 310282, 310283, 310284, 310285, 310286, 310287, 310288, 310289, 310290, 310291, 310292, 310293, 310294, 310295, 310296, 310297, 310298, 310299, 310300, 310301, 310302, 310303, 310304, 310305, 310306, 310307, 310308, 310309, 310310, 310311, 310313, 310315, 310316, 310317, 310318, 310319, 310320, 310321, 310322, 310323, 310324, 310325, 310326, 310327, 310328, 310329, 310330, 310331, 333002, 333005, 333009, 333010, 333014, 333015, 333018, 333019, 333021, 333030, 333031, 333037, 333041, 333042, 333043, 333044, 333045, 333046, 333048, 333049, 333050, 333051, 333052, 333053, 333054, 333055, 333056, 333059, 333060, 333061, 333062, 333063, 333064, 333065, 333066, 333067, 333068, 333069, 333070, 333071, 333072, 333073, 333074, 333075, 333076, 333077, 333078, 333079, 333080, 333081, 333082, 333083, 333085, 333086, 333087, 333088, 333089, 333090, 333091, 333092, 333093, 333094, 333095, 333096, 333097, 333098, 333099, 333100, 333101, 333102, 333103, 333104, 333105, 333106, 333107, 333108, 333109, 333110, 333111, 333112, 333113, 333114, 333115, 333116, 333117, 333118, 333119, 333120, 333121, 333122, 333123, 333124, 333125, 333126, 333127, 333129, 333130, 333131, 333132, 333133, 333134, 333135, 333136, 333137, 333140, 333144, 333145, 333146, 333150, 333151

Distribution pattern

Worldwide Distribution - US Nationwide in the states of:: AK, AL, AR, AZ, CA, CO, CT, DE, GA, IA, IL, IN, FL, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NJ, NH, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, & WY. and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Brazil, Cambodia, China, COlumbia, Cuba, Denmark, Dijbouti, Dominican REpublic, Egypt, Estonia, Finland, France, Germany, Gilbraltar, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Libya, Lithuania, Malaysia, Martinique, Monaco, Mongolia, Mozambique, Myanmar, Netherlands, New Zealand, Norway, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russian Federation, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Seitzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uniter Arab Emirates, United Kingdom, Uzbekistan, and Vietnam.

device · product 3 of 4

Ingenuity Core 128 Model Number 728323 Product Usage: Computed Tomography X-ray systems intended to product cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Z-0011-2017
Recall number
Z-0011-2017
Initiated
August 24, 2016
Classification
Class II
Status
Terminated
Recalling firm
Philips Medical Systems
Quantity
524

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults to standard resolution which doesn't experience the issue.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults to standard resolution which doesn't experience the issue.

Code information

30016, 30023, 30024, 30151, 32000, 32001, 32008, 32010, 32012, 32013, 32015, 32016, 32018, 32021, 32022, 32024, 32026, 32028, 32032, 32033, 32034, 32036, 32037, 32041, 32042, 32043, 32044, 32046, 32047, 32048, 32049, 32050, 32051, 32052, 32053, 32054, 32055, 32056, 32057, 32058, 32059, 32060, 32061, 32062, 32063, 32064, 32065, 32066, 32067, 32068, 32069, 32070, 32071, 32072, 32073, 32074, 32075, 32076, 32077, 32078, 32079, 32080, 32081, 32082, 32084, 32085, 32086, 32087, 32088, 32089, 32090, 32091, 32092, 32093, 32094, 32095, 32096, 32097, 32098, 32099, 32100, 32101, 32102, 32103, 32104, 32105, 32106, 32108, 32109, 32110, 32111, 32112, 32113, 32114, 32115, 32116, 32117, 32118, 32119, 32120, 32121, 32122, 32123, 32124, 32125, 32126, 32127, 32128, 32129, 32130, 32131, 32133, 32134, 32135, 32136, 32137, 32138, 32139, 32140, 32141, 32142, 32143, 32144, 32145, 32146, 32147, 32148, 32149, 32150, 32151, 32152, 32153, 32154, 32155, 32156, 32157, 32158, 32159, 32160, 32161, 32162, 32163, 32164, 32165, 32166, 32167, 32168, 32169, 32171, 32173, 52023, 52028, 52049, 52073, 320001, 320002, 320003, 320005, 320006, 320008, 320010, 320018, 320019, 320024, 320026, 320029, 320030, 320031, 320033, 320037, 320054, 320058, 320059, 320069, 320073, 320074, 320078, 320079, 320080, 320081, 320082, 320083, 320084, 320088, 320090, 320092, 320096, 320099, 320100, 320101, 320103, 320107, 320111, 320114, 320115, 320119, 320122, 320123, 320125, 320126, 320128, 320129, 320131, 320132, 320133, 320134, 320136, 320138, 320144, 320146, 320147, 320200, 320201, 320202, 320203, 320204, 320206, 320207, 320208, 320209, 320210, 320211, 320212, 320213, 320214, 320215, 320216, 320217, 320218, 320219, 320220, 320221, 320222, 320223, 320224, 320225, 320226, 320227, 320228, 320229, 320230, 320231, 320232, 320233, 320234, 320235, 320236, 320237, 320238, 320239, 320240, 320241, 320242, 320243, 320245, 320246, 320247, 320248, 320250, 320251, 320252, 320253, 320254, 320255, 320256, 320257, 320258, 320259, 320260, 320261, 320262, 320263, 320264, 320265, 320266, 320267, 320268, 320269, 320270, 320271, 320272, 320273, 320274, 320275, 320276, 320278, 320279, 320280, 320281, 320282, 320283, 320284, 320285, 320286, 320288, 320289, 320290, 320291, 320292, 320293, 320295, 320296, 320297, 320298, 320299, 320301, 320302, 320303, 320305, 320307, 320309, 320310, 320311, 320312, 320313, 336011, 336012, 336013, 336015, 336016, 336017, 336018, 336023, 336025, 336027, 336028, 336029, 336034, 336040, 336044, 336046, 336051, 336059, 336060, 336064, 336066, 336067, 336068, 336069, 336070, 336071, 336072, 336073, 336075, 336076, 336077, 336078, 336079, 336080, 336081, 336082, 336083, 336084, 336085, 336086, 336087, 336088, 336089, 336090, 336091, 336092, 336093, 336094, 336095, 336096, 336097, 336098, 336099, 336100, 336101, 336102, 336103, 336104, 336105, 336106, 336107, 336108, 336109, 336110, 336111, 336112, 336113, 336114, 336115, 336116, 336117, 336118, 336119, 336120, 336122, 336123, 336124, 336125, 336126, 336127, 336128, 336129, 336130, 336131, 336132, 336133, 336134, 336135, 336136, 336137, 336138, 336139, 336140, 336141, 336142, 336143, 336144, 336145, 336146, 336147, 336148, 336149, 336150, 336151, 336152, 336153, 336154, 336155, 336156, 336157, 336158, 336159, 336160, 336161, 336162, 336163, 336164, 336165, 336166, 336167, 336168, 336169, 336170, 336171, 336172, 336173, 336174, 336175, 336176, 336177, 336178, 336179, 336180, 336181, 336182, 336183, 336184, 336185, 336186, 336187, 336188, 336189, 336190, 336191, 336192, 336193, 336194, 336195, 336196, 336197, 336198, 336199, 336200, 336201, 336202, 336203, 336204, 336205, 336206, 336207, 336208, 336209, 336210, 336211, 336212, 336213, 336214, 336215, 336216, 336217, 336218, 336219, 336220, 336221, 336222, 336223, 336224, 336225, 336226, 336227, 336228, 336229, 336230, 336231, 336232, 336234, 336235, 336236, 336241, 336243, 336245, 336246, 336247, 336248, 336249, 336252, 336256, 336781, 336782, 336783, 336784, 336785, 336786, 336787, 336788, 336789, 336790, 336791

Distribution pattern

Worldwide Distribution - US Nationwide in the states of:: AK, AL, AR, AZ, CA, CO, CT, DE, GA, IA, IL, IN, FL, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NJ, NH, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, & WY. and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Brazil, Cambodia, China, COlumbia, Cuba, Denmark, Dijbouti, Dominican REpublic, Egypt, Estonia, Finland, France, Germany, Gilbraltar, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Libya, Lithuania, Malaysia, Martinique, Monaco, Mongolia, Mozambique, Myanmar, Netherlands, New Zealand, Norway, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russian Federation, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Seitzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uniter Arab Emirates, United Kingdom, Uzbekistan, and Vietnam.

device · product 4 of 4

Ingenuity CT Model Number 728326 Product Usage: Computed Tomography X-ray systems intended to product cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Z-0012-2017
Recall number
Z-0012-2017
Initiated
August 24, 2016
Classification
Class II
Status
Terminated
Recalling firm
Philips Medical Systems
Quantity
235

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults to standard resolution which doesn't experience the issue.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults to standard resolution which doesn't experience the issue.

Code information

30003, 30004, 30005, 30006, 30007, 30008, 30009, 30010, 30011, 30012, 30013, 30014, 30015, 30017, 30018, 30019, 30020, 30021, 30022, 30025, 30026, 30027, 30028, 30029, 30030, 30031, 30032, 30033, 30034, 30036, 30037, 30038, 30039, 30040, 30041, 30042, 30043, 30044, 30045, 30046, 30047, 30049, 32045, 32083, 32107, 52051, 52060, 52062, 300003, 300004, 300005, 300009, 300010, 300011, 300012, 300013, 300014, 300015, 300016, 300017, 300018, 300019, 300020, 300021, 300022, 300024, 300025, 300027, 300028, 300029, 300030, 300031, 300032, 300033, 300034, 300035, 300036, 300037, 300038, 300039, 300040, 300041, 300042, 300043, 300044, 300045, 300047, 300049, 300050, 300051, 300052, 300053, 300054, 300055, 300056, 300057, 300058, 300059, 300060, 300061, 300062, 300063, 300064, 300065, 300066, 300068, 300069, 300070, 300071, 300072, 300073, 300074, 300075, 300077, 300078, 300079, 300080, 300081, 300082, 300083, 300084, 300086, 300087, 300088, 300089, 300090, 300091, 300092, 300093, 300094, 300095, 300096, 300097, 300098, 300099, 300100, 300101, 300102, 300103, 300104, 300105, 300106, 300107, 300108, 300109, 300110, 300111, 300112, 300113, 300114, 300115, 300116, 300117, 300118, 300119, 300120, 300121, 300122, 300123, 300124, 300125, 300126, 300127, 300128, 300129, 300130, 300131, 300132, 300133, 300135, 300136, 300137, 300138, 300139, 300140, 300141, 300142, 300143, 300144, 300145, 300147, 300148, 300149, 300150, 300151, 300156, 300157, 300158, 300159, 300160, 300161, 300162, 300163, 300164, 300165, 300166, 300200, 300201, 300202, 300203, 300204, 300205, 300206, 300207, 300208, 300209, 300210, 300211, 300212, 300213, 300214, 300146, 600023, 600047, 300215, 300216, 300217, 300218, 300219, 300220, 300221, 300223, 310058, 310144, 320205, 333057, 333084, 333128, 336061, 336084, 336121, 336233, 30004-1

Distribution pattern

Worldwide Distribution - US Nationwide in the states of:: AK, AL, AR, AZ, CA, CO, CT, DE, GA, IA, IL, IN, FL, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NJ, NH, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, & WY. and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Brazil, Cambodia, China, COlumbia, Cuba, Denmark, Dijbouti, Dominican REpublic, Egypt, Estonia, Finland, France, Germany, Gilbraltar, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Libya, Lithuania, Malaysia, Martinique, Monaco, Mongolia, Mozambique, Myanmar, Netherlands, New Zealand, Norway, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russian Federation, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Seitzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uniter Arab Emirates, United Kingdom, Uzbekistan, and Vietnam.