Event 75167

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 07, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Ortho-Clinical Diagnostics

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

enGen Laboratory Automation System (Product Code ENGEN) configured with Thermo Scientific Centrifuge Module and TCAutomation Software Version 3.6.1 and Below; IVD. --- Thermo Scientific Centrifuge Module has Product Code 952040-EG 6844097. The VITROS 5,1 FS Chemistry System with enGen Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods).

Z-0300-2017
Recall number
Z-0300-2017
Initiated
September 07, 2016
Classification
Class II
Status
Terminated
Quantity
60 (US: 24; Foreign: 36)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Ortho confirmed that it is possible for the mis-association of sample identification and assay results due to an error with Thermo Fishers centrifuge module. It is possible for mis-associated assay results to be reported from the laboratory prior to the identification of the Cross Check error on the enGen Laboratory Automation System. Assay results associated with the wrong patient and reported out of the laboratory could lead to inappropriate intervention with the potential for serious injury to the patient.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Ortho confirmed that it is possible for the mis-association of sample identification and assay results due to an error with Thermo Fishers centrifuge module. It is possible for mis-associated assay results to be reported from the laboratory prior to the identification of the Cross Check error on the enGen Laboratory Automation System. Assay results associated with the wrong patient and reported out of the laboratory could lead to inappropriate intervention with the potential for serious injury to the patient.

Code information

Software Versions 3.6.1 and below; Affected US Serial Numbers: 5GV6W2J,GBX395J, J2QKG25J, 2Z3YV1J, 3RHC23J, 4HX7S4J, 54FC01J, 5J4B45J, 736BM4J, 7KSC15J, J86PWC5J, 953209, 953221, 953266, 953270, B3ZHP2J, CXL941J, JFBX395J, FCM073J, FXCBG3J, JG7BFB5J, GWXQ52J, HJ5K84J, HPKG25J -- Affected Foreign Serial Numbers: DWXO52J, GWJZY4J, J16XQH4J, BGV6W2J, 17QP94J, 4SHPF4J, 1K5K84J, 37QP34J, H5PWC5J, J2V6LW4J, F6MXR3J, JT6LW4J, N0551107, F1150698, 46XQH4J, 2S5OP0J, CRKM82, J41BCY4J, J56T205J, BGV6W2J, 7YCF43J, 96XQHAJ, C1JNZ5J, C3JTP4J, D3JTP4J, 86T205J, 18BFB5J, 96PWC5J, 1LSSF5J, 4drj44j, 8W7WG3J, J953222, J953244, JCT20Q3J, HWJZY4J, 38769917923.

Distribution pattern

Worldwide Distribution-(Nationwide) and foreign distribution to the following countries: Belgium, Bermuda, Canada, Chile, Denmark, France, Italy, Mexico, Norway, Portugal, Sweden, Taiwan and United Kingdom.