Timeline bucket
August 09, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Hill-Rom, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling installed rail systems: Multirall 200, Likorall 200, 242, 243, 250, and LikoGuard L and XL.

Z-0326-2017
Recall number
Z-0326-2017
Initiated
August 09, 2016
Classification
Class II
Status
Terminated
Recalling firm
Hill-Rom, Inc.
Quantity
188

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Ceiling Bracket 71 was shipped with a larger center hole. A potentially hazardous situation exists that if the Ceiling Bracket 71 with a larger center hole is used for a concrete ceiling Ultra System installation with the M8 screw and the safe working load is exceeded, the system could fall.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Mixed-up of materials/components

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Ceiling Bracket 71 was shipped with a larger center hole. A potentially hazardous situation exists that if the Ceiling Bracket 71 with a larger center hole is used for a concrete ceiling Ultra System installation with the M8 screw and the safe working load is exceeded, the system could fall.

Code information

manufacturing date between Dec 01, 2015- Apr 14, 2016

Distribution pattern

WA NC OH PA MI Australia Sweden