Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 75255

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 14, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Hand Biomechanics Lab Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Agee-WristJack¿ Fracture Reduction System; Item number: CFD-147. The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. This is a single use system used for a trauma related surgery The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. The non-perishable sterile system is packaged in an inner then outer tray and shipped in a cardboard shelf box. Orthopedic:The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures.

Z-0136-2017
Recall number
Z-0136-2017
Initiated
September 14, 2016
Classification
Class II
Status
Terminated
Recalling firm
Hand Biomechanics Lab Inc
Quantity
65 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A number of Agee WristJack¿ Fracture Reduction Systems have been manufactured with a molded beam slightly larger than optimal. This may result in the surgeon experiencing difficulty fine-tuning the length (red) and radial/ulnar alignment (blue) adjustments on the device.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A number of Agee WristJack¿ Fracture Reduction Systems have been manufactured with a molded beam slightly larger than optimal. This may result in the surgeon experiencing difficulty fine-tuning the length (red) and radial/ulnar alignment (blue) adjustments on the device.

Code information

Lot number: HBL-115-165, serial numbers 47460959 and 47460980; Lot number: HBL-116-014 - all serial numbers; Lot number : HBL-116-040 - 15 systems shipped prior to recall

Distribution pattern

Nationwide Distribution