Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 75306

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 15, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Pentax of America Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Pentax Video Colonoscope is intended to be used with a PENTAX video processor (including light source), documentation equipment, Monitor, EndoTherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract including the anus, rectum, sigmoid colon, colon and ileocecal valve.

Z-0636-2017
Recall number
Z-0636-2017
Initiated
August 15, 2016
Classification
Class II
Status
Terminated
Recalling firm
Pentax of America Inc
Quantity
5,090 (US) and 14,209 (OUS)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Pentax is initiating the product correction of several model numbers of Pentax's Colonoscope that were shipped between the clearance date of April 2014 and September 2015 because they contain the operation and reprocessing IFU's that were in use prior to the 2014 clearances and therefore did not include the cleared operation IFU, the updated reprocessing IFU or the dedicated reprocessing adaptor (for applicable models).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Other

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Pentax is initiating the product correction of several model numbers of Pentax's Colonoscope that were shipped between the clearance date of April 2014 and September 2015 because they contain the operation and reprocessing IFU's that were in use prior to the 2014 clearances and therefore did not include the cleared operation IFU, the updated reprocessing IFU or the dedicated reprocessing adaptor (for applicable models).

Code information

Model #'s EC-2990LI, EC-3490LI, EC-3490LK, EC-3490TLI, EC-3890LI, EC-3890LK and EC-3890TLK

Distribution pattern

US Nationwide and Japan