Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 75513

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 04, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Inpeco S.A.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

FlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

Z-0587-2017
Recall number
Z-0587-2017
Initiated
October 04, 2016
Classification
Class II
Status
Terminated
Recalling firm
Inpeco S.A.
Quantity
188 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Inpeco received several notifications from the field about burnt schuko sockets. The customers reported that some smoke was seen coming from the damaged sockets.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Employee error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Inpeco received several notifications from the field about burnt schuko sockets. The customers reported that some smoke was seen coming from the damaged sockets.

Code information

Shucko Socket with P/N: 693822160, Model: CMSA/HETTICH - STORAGE - i2000SR, Length: 3 meters; Shucko Socket with P/N: 693822170, Model: CMSA/HETTICH - STORAGE - i2000SR, Length: 5 meters; Shucko Socket with P/N: 693822190, Model: CMSA/HETTICH - STORAGE - i2000SR, Length: 7 meters; Shucko Socket with P/N: 0C00005640, Model: CMSA/HETTICH - STORAGE - i2000SR, Length: 2 meters. The Serial number of the impacted products globally distributed are the following: ACP.009, ACP.024, ACP.039, ACP.054, ACP.069, ACP.084, ACP.100, ACP.115 , ACP.010, ACP.025, ACP.040, ACP.055, ACP.070, ACP.085, ACP.101, ACP.116, ACP.011, ACP.026, ACP.041, ACP.056, ACP.071, ACP.086, ACP.102, ACP.117, ACP.012, ACP.027, ACP.042, ACP.057, ACP.072,ACP.087, ACP.103, ACP.118 , ACP.013, ACP.028, ACP.043, ACP.058, ACP.073, ACP.088, ACP.104, ACP.119, ACP.014, ACP.029, ACP.044, ACP.059, ACP.074, ACP.089, ACP.105, ACP.120, ACP.015, ACP.030, ACP.045, ACP.060, ACP.075, ACP.090, ACP.110, ACP.121, ACP.016, ACP.031, ACP.046, ACP.061, ACP.076, ACP.091, ACP.107, ACP.122 ,ACP.017, ACP.032, ACP.047, ACP.062, ACP.077, ACP.092, ACP.108, ACP.123 ,ACP.018, ACP.033, ACP.048, ACP.063, ACP.078, ACP.093, ACP.109, ACP.124, ACP.019, ACP.034, ACP.049, ACP.064, ACP.079, ACP.094, ACP.110, ACP.125, ACP.020, ACP.035, ACP.050, ACP.065, ACP.080, ACP.096, ACP.111, ACP.126 ,ACP.021, ACP.036, ACP.051, ACP.066, ACP.081, ACP.097, ACP.112, ACP.127 , ACP.022, ACP.037, ACP.052, ACP.067, ACP.082, ACP.098, ACP.113, ACP.128, ACP.023, ACP.038, ACP.053, ACP.068, ACP.083, ACP.099, ACP.114, ACP.129, FLX.103, FLX.0120, FLX.0131, FLX.0137, FLX.0143, FLX.0161, FLX.0166, FLX.0179, FLX.0104, FLX.0124, FLX.0134, FLX.0140, FLX.0152, FLX.0163, FLX.0175, FLX.0108, FLX.0125, FLX.0135, FLX.0141 FLX.0155, FLX.0164, FLX.0176, FLX.0115, FLX.0126, FLX.0136, FLX.0142, FLX.0158, FLX.0165, FLX.0178, AP2.0001, AP2.0004, AP2.0006, AP2.0008, AP2.0010, AP2.0012, AP2.0014, AP2.0003, AP2.0002, AP2.0005, AP2.0007, AP2.0009, AP2.0011, AP2.0013, AP2.0015, FLX.0008, FLX.0133, FLX.0149, FLX.0153, FLX.0159, FLX.0167, FLX.0170, FLX.0174, FLX.0069, FLX.0139, FLX.0150 FLX.0156, FLX.0162, FLX.0168, FLX.0172, FLX.0179, FLX.0132, FLX.0147, FLX.0151, FLX.0157, FLX.0166, FLX.0169, FLX.0173 ,FLX.0121, FLX.0138, FLX.0144, FLX.0145, FLX.0146, FLX.0148, FLX.0176 and FLX.0181.

Distribution pattern

Worldwide Distribution - US including Texas and New York. Foreign Consignees: Germany, Italy, France, Austria and Spain.