Monaco TRP System The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy.
- Recall number
- Z-0668-2017
- Initiated
- November 17, 2016
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Elekta, Inc.
- Quantity
- 2148
App-derived interpretation
Contours are not handled correctly if the CT dataset contains slices with DICOM Z coordinates of xx.x50mm.
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
Contours are not handled correctly if the CT dataset contains slices with DICOM Z coordinates of xx.x50mm.
Code information
Software Builds 5.00 and higher
Distribution pattern
Worldwide Distribution - US including AK, AZ, CA, CO, FL, GA, IL, IN, KY, LA, MD, MA, MI, MN, MO, MT, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, SD, TX, UT, VA, WA, WI, Puerto Rico and Internationally to Algeria, Antiqua & Barbuda, Argentina, Australia, Austria, Bahrain, Belarus, Botswana, Brazil, Bulgaria, Canada, Chile, China, Columbia, Croatia, Cuba, Cyprus, Czech Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Laos, Lithuania, Malaysia, Malta, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Norway, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Thailand, Tunisia, Turkey, Unite Kingdom, Ukraine, Venezuela, Vietnam, and Zimbabwe.