device · product 1 of 1
Tosoh Bioscience HLC-723G8 Analyzer for HbA1c testing.
- Recall number
- Z-1225-2017
- Initiated
- November 07, 2016
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Tosoh Bioscience, Inc.
- Quantity
- 682 analyzers
App-derived interpretation
The device defect associated with this recall may result in HbA1c values with a negative bias in patients with the hemoglobin variants C, D, E and S. The negative bias occurs at the higher levels of HbA1c, e.g. 9%, but not at the diagnostic medical decision point of 6.5%.
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
The device defect associated with this recall may result in HbA1c values with a negative bias in patients with the hemoglobin variants C, D, E and S. The negative bias occurs at the higher levels of HbA1c, e.g. 9%, but not at the diagnostic medical decision point of 6.5%.
Code information
All active analyzers
Distribution pattern
Worldwide distribution. The products were shipped to the following states: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, AND WY. The recalled products were shipped to the following countries: Brazil, Chile, Columbia, Guatemala, Costa Rica, Dominican Republic, Eduador, Nicaragua, Panama, Peru, Puerto Rico, Uruguay,