Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 75860

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 22, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Medical Solutions USA, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

SOMATOM Scope, Computed tomography x-ray system

Z-0919-2017
Recall number
Z-0919-2017
Initiated
November 22, 2016
Classification
Class II
Status
Terminated
Quantity
13 systems

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.

Code information

Model # 10967666 Serial Number:59376 91766 91767 115456 115454 59380 77560 98840 77557 91758 59377 59378 115451

Distribution pattern

Distributed to: IL IL GA SC SD OK NY MI IN MD AR

device · product 2 of 3

SOMATOM Scope Power, Computed tomography x-ray system

Z-0920-2017
Recall number
Z-0920-2017
Initiated
November 22, 2016
Classification
Class II
Status
Terminated
Quantity
N/A

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.

Code information

Model # 10967888 Serial Number:59376 91766 91767 115456 115454 59380 77560 98840 77557 91758 59377 59378 115451

Distribution pattern

Distributed to: IL IL GA SC SD OK NY MI IN MD AR

device · product 3 of 3

SOMATOM Perspective, Computed tomography x-ray system

Z-0921-2017
Recall number
Z-0921-2017
Initiated
November 22, 2016
Classification
Class II
Status
Terminated
Quantity
13 systems

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.

Code information

Model # 10967888 Serial Number:59376 91766 91767 115456 115454 59380 77560 98840 77557 91758 59377 59378 115451

Distribution pattern

Distributed to: IL IL GA SC SD OK NY MI IN MD AR