Event summary

Timeline bucket
November 22, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Interventional Spine Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

G3 Opticage Expandable Interbody Fusion Device, Model No. 9180-00, 9180-01, 9180-06, 9180-07, 9180-17, 9101-00, 9101-06 The Opticage Expandable Interbody Fusion Device is a posterior lumbar intervertebral body fusion device and is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Opticage Interbody Fusion Device can be implanted via posterior, transforaminal or lateral approach. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Candidates for surgery should be skeletally mature and have had six months of conservative treatment. These patients may have had primary or secondary surgery, but no previous fusion at the involved levels. The device is not intended to be used as a stand-alone device. It must be used with supplemental internal spinal fixation systems that have been cleared for use in the lumbar spine (i.e. facet screw fixation systems, facet compression devices and posterior pedicle screw and rod systems).

Z-0941-2017
Recall number
Z-0941-2017
Initiated
November 22, 2016
Classification
Class II
Status
Terminated
Recalling firm
Interventional Spine Inc
Quantity
239 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
During implantation, the proximal end of the device could separate if the angle between the device and insertion instrument is too steep. This separation will prevent the device from expanding.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

During implantation, the proximal end of the device could separate if the angle between the device and insertion instrument is too steep. This separation will prevent the device from expanding.

Code information

Lot no. 011416-D 021216-A 021216-B 032015-B 032015-C 033116-B 040915-B 040915-E 041416-A 060116-A 060116-B 060116-C 061616-B 061915-A 061915-B 070714-D 071116-A 071116-B 071116-C 071516-A 071516-B 071916-A 081016-A 081016-B 081016-C 083116-B 083116-B 091516-A 112515-A 033116-C 030716-A 081016-E 090216-C 090216-C 090216-D 030716-B 011416-E 021216-C 081016-F 081016-G 090216-A 090216-A 090216-B 081016-H 083116-C 083116-C 083116-D

Distribution pattern

Nationwide Distribution to NY, NC, SC, PA, TX, OH, GA, NJ, DE, AZ, DC, CA, and VA.