Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 76018

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 02, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Varian Medical Systems, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Eclipse Treatment Planning System version 13.MR2 [13.06.31 with Smart Segmentation Knowledge Based Contouring version 2.3 [2.3.12] Radiology: The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments.

Z-0990-2017
Recall number
Z-0990-2017
Initiated
December 02, 2016
Classification
Class II
Status
Terminated
Quantity
42 devices are affected

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Modifications in version 13.6MR2 for Contouring, SmartAdapt, and SmartSegmentation workspaces resulted in contours not being saved consistently in Eclipse. Treatment Planning System. The issue only occurs if certain conditions are fulfilled.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Modifications in version 13.6MR2 for Contouring, SmartAdapt, and SmartSegmentation workspaces resulted in contours not being saved consistently in Eclipse. Treatment Planning System. The issue only occurs if certain conditions are fulfilled.

Code information

Model # H48 Serial numbers: H48L574 H48L575 H48L576 H48L577 H48L578 H48L579 H48L580 H48L581 H48L582 H48L583 H48L584 H48L585 H48L586 H48L587 H482484 H487302 H48A587 H48D957 H480113 H480118 H481884 H481885 H481886 H481916 H481917 H484198 H484199 H488949 H48A680 H48A871 H48A872 H48A873 H48A874 H48A875 H48A876 H48A877 H48A878 H48A879 H48I531 H481195 H481236 H481237.

Distribution pattern

US Distribution to the states of : NJ, TN. OR and FL.