device · product 1 of 1
JResultNet drivers using ASTM coding language. Calculator/data processing module, for clinical use - Stand-alone software package that provides connectivity between in-vitro diagnostic instruments and laboratory information's systems, aiding the flow of laboratory information, including test orders and results, specimen information, and patient information
- Recall number
- Z-1680-2017
- Initiated
- October 23, 2014
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Data Innovations, LLC
- Quantity
- 657 units
App-derived interpretation
Software issue
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
Software issue resulting in the results of one patient potentially being associated with another.
Code information
All drive versions were affected.
Distribution pattern
Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, & DC.