device · product 1 of 1
ProSpore 4 mL Ampoule Biological indicator for steam sterilization, contains Geobacillus stearothermophilus, Catalog Number: PS-6-50, contains 50 Ampoules/Box. Product Usage: Biological indicator for steam sterilization at 121 deg C intended for use by a health care provider to accompany products being sterilized through steam sterilization procedure and to monitor adequacy of sterilization.
Product resistance has fallen outside of label claim within expiry.
Code information
Lot # 637, exp. date 12/2016
Distribution pattern
Product was distributed to: AL, AR, CA, FL, IA, MO, NE, NC, NY, OH, and WA. The product was also distributed internationally, including Mexico and Canada. There was no goverment or military distribution.