Event 76130

11 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 28, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Medtronic Advanced Energy, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

11 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 11

Intact Disposable Wands (part of the Intact System) 777-110: 5 Pack 10mm Disposable Product Usage: The Intact Medical disposable wands are intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality and to provide breast tissue for histologic examination with partial removal of a palpable abnormality.

Z-1126-2017
Recall number
Z-1126-2017
Initiated
December 28, 2016
Classification
Class II
Status
Terminated
Quantity
10 units

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
sterility compromised

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Product sterility compromised due to breach of sterile barrier

Code information

All lot codes

Distribution pattern

Worldwide Distribution - US Nationwide in the countries of China, Greece, Netherlands, Egypt, Turkey, and Switzerland.

device · product 2 of 11

Intact Disposable Wands (part of the Intact System) 900-112: 5 Pack 12mm Disposable Product Usage: The Intact Medical disposable wands are intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality and to provide breast tissue for histologic examination with partial removal of a palpable abnormality.

Z-1127-2017
Recall number
Z-1127-2017
Initiated
December 28, 2016
Classification
Class II
Status
Terminated
Quantity
155 units

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
sterility compromised

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Product sterility compromised due to breach of sterile barrier

Code information

All lot codes

Distribution pattern

Worldwide Distribution - US Nationwide in the countries of China, Greece, Netherlands, Egypt, Turkey, and Switzerland.

device · product 3 of 11

Intact Disposable Wands (part of the Intact System) 777-112: 5 Pack 12mm Disposable Product Usage: The Intact Medical disposable wands are intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality and to provide breast tissue for histologic examination with partial removal of a palpable abnormality.

Z-1128-2017
Recall number
Z-1128-2017
Initiated
December 28, 2016
Classification
Class II
Status
Terminated
Quantity
180 units

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
sterility compromised

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Product sterility compromised due to breach of sterile barrier

Code information

All lot codes

Distribution pattern

Worldwide Distribution - US Nationwide in the countries of China, Greece, Netherlands, Egypt, Turkey, and Switzerland.

device · product 4 of 11

Intact Disposable Wands (part of the Intact System) 777-112BT: 5 Pack 12mm Disposable Product Usage: The Intact Medical disposable wands are intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality and to provide breast tissue for histologic examination with partial removal of a palpable abnormality.

Z-1129-2017
Recall number
Z-1129-2017
Initiated
December 28, 2016
Classification
Class II
Status
Terminated
Quantity
252 units

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
sterility compromised

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Product sterility compromised due to breach of sterile barrier

Code information

All lot codes

Distribution pattern

Worldwide Distribution - US Nationwide in the countries of China, Greece, Netherlands, Egypt, Turkey, and Switzerland.

device · product 5 of 11

Intact Disposable Wands (part of the Intact System) 900-115: 5 Pack 15mm Disposable Product Usage: The Intact Medical disposable wands are intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality and to provide breast tissue for histologic examination with partial removal of a palpable abnormality.

Z-1130-2017
Recall number
Z-1130-2017
Initiated
December 28, 2016
Classification
Class II
Status
Terminated
Quantity
185 units

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
sterility compromised

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Product sterility compromised due to breach of sterile barrier

Code information

All lot codes

Distribution pattern

Worldwide Distribution - US Nationwide in the countries of China, Greece, Netherlands, Egypt, Turkey, and Switzerland.

device · product 6 of 11

Intact Disposable Wands (part of the Intact System) 777-115: 5 Pack 15mm Disposable Product Usage: The Intact Medical disposable wands are intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality and to provide breast tissue for histologic examination with partial removal of a palpable abnormality.

Z-1131-2017
Recall number
Z-1131-2017
Initiated
December 28, 2016
Classification
Class II
Status
Terminated
Quantity
320 units

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
sterility compromised

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Product sterility compromised due to breach of sterile barrier

Code information

All lot codes

Distribution pattern

Worldwide Distribution - US Nationwide in the countries of China, Greece, Netherlands, Egypt, Turkey, and Switzerland.

device · product 7 of 11

Intact Disposable Wands (part of the Intact System) 777-115BT: 5 Pack 15mm Disposable Product Usage: The Intact Medical disposable wands are intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality and to provide breast tissue for histologic examination with partial removal of a palpable abnormality.

Z-1132-2017
Recall number
Z-1132-2017
Initiated
December 28, 2016
Classification
Class II
Status
Terminated
Quantity
499 units

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
sterility compromised

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Product sterility compromised due to breach of sterile barrier

Code information

All lot codes

Distribution pattern

Worldwide Distribution - US Nationwide in the countries of China, Greece, Netherlands, Egypt, Turkey, and Switzerland.

device · product 8 of 11

Intact Disposable Wands (part of the Intact System) 900-120: 5 Pack 20mm Disposable Product Usage: The Intact Medical disposable wands are intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality and to provide breast tissue for histologic examination with partial removal of a palpable abnormality.

Z-1133-2017
Recall number
Z-1133-2017
Initiated
December 28, 2016
Classification
Class II
Status
Terminated
Quantity
275 units

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
sterility compromised

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Product sterility compromised due to breach of sterile barrier

Code information

All lot codes

Distribution pattern

Worldwide Distribution - US Nationwide in the countries of China, Greece, Netherlands, Egypt, Turkey, and Switzerland.

device · product 9 of 11

Intact Disposable Wands (part of the Intact System) 777-120: 5 Pack 20mm Disposable Product Usage: The Intact Medical disposable wands are intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality and to provide breast tissue for histologic examination with partial removal of a palpable abnormality.

Z-1134-2017
Recall number
Z-1134-2017
Initiated
December 28, 2016
Classification
Class II
Status
Terminated
Quantity
326 units

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
sterility compromised

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Product sterility compromised due to breach of sterile barrier

Code information

All lot codes

Distribution pattern

Worldwide Distribution - US Nationwide in the countries of China, Greece, Netherlands, Egypt, Turkey, and Switzerland.

device · product 10 of 11

Intact Disposable Wands (part of the Intact System) 777-120BT: 5 Pack 20mm Disposable Product Usage: The Intact Medical disposable wands are intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality and to provide breast tissue for histologic examination with partial removal of a palpable abnormality.

Z-1135-2017
Recall number
Z-1135-2017
Initiated
December 28, 2016
Classification
Class II
Status
Terminated
Quantity
996 units

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
sterility compromised

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Product sterility compromised due to breach of sterile barrier

Code information

All lot codes

Distribution pattern

Worldwide Distribution - US Nationwide in the countries of China, Greece, Netherlands, Egypt, Turkey, and Switzerland.

device · product 11 of 11

Intact Disposable Wands (part of the Intact System) 900-130: 5 Pack 30mm Disposable Product Usage: The Intact Medical disposable wands are intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality and to provide breast tissue for histologic examination with partial removal of a palpable abnormality.

Z-1136-2017
Recall number
Z-1136-2017
Initiated
December 28, 2016
Classification
Class II
Status
Terminated
Quantity
105 units

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
sterility compromised

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Product sterility compromised due to breach of sterile barrier

Code information

All lot codes

Distribution pattern

Worldwide Distribution - US Nationwide in the countries of China, Greece, Netherlands, Egypt, Turkey, and Switzerland.