device · product 1 of 1
Carestream DRX-Revolution Mobile X-Ray System, Model DRXR-1, Catalog No. 1019397 (all other countries), Catalog No. 1060177 (China only) -- Carestream Health, Inc. 150 Verona Street, Rochester, NY 14608
- Recall number
- Z-1138-2017
- Initiated
- January 12, 2017
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Carestream Health Inc
- Quantity
- US: 1403 units; Foreign: 1370 units
App-derived interpretation
The firm received a complaint alleging that a DRX Revolution System could not be stopped when applying the brake. The Revolution crashed into an elevator.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
The firm received a complaint alleging that a DRX Revolution System could not be stopped when applying the brake. The Revolution crashed into an elevator.
Code information
Serial Numbers: 101 to 2545 (Worldwide except China), 800101 to 800468 (China only)
Distribution pattern
Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico, and to the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Denmark, Dominican Republic, Dubai, Finland, France, Germany, Ghana, India, Iran, Ireland, Israel, Italy, Japan, Korea, Kuwait, Lebanon, Libya, Malaysia, Maldives, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates and United Kingdom.