Event 76275

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 13, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Vertebral Technologies, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

InterFuse Tail Traction Tool, Part Number 9116. The InterFuse Tail Traction Tool is an optional Class I non-sterile reusable instrument that can be used to assist with module locking of the InterFuse S or InterFuse T implant system. The system has a modular design to allow the device to be assembled within the disc space to minimize the surgical window, while providing a footprint that approximates the excised nucleus.

Z-1151-2017
Recall number
Z-1151-2017
Initiated
January 13, 2017
Classification
Class II
Status
Terminated
Quantity
63 pieces

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Fit issues have been identified with the Tail Traction Tool (PN 9116) when used with the Interfuse T tailed implant. The InterFuse T Tail may not fit into the Thru Hole in the instrument.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Fit issues have been identified with the Tail Traction Tool (PN 9116) when used with the Interfuse T tailed implant. The InterFuse T Tail may not fit into the Thru Hole in the instrument.

Code information

Lot: 110512-02, 110718-01, 110831-02, 120214-01, 120403-018, 120420-07, 120711-02, 120911-06

Distribution pattern

US: IL, WI, OH, NY, MT, FL, MD, IN, TN, TX, MN. OUS: Italy, Taiwan Spain, Belgium, Canada, Netherlands, Sweden, Switzerland, Portugal, Brazil. Germany, Ireland,