Event 76656

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 06, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Electronics North America Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Philips IntelliVue MX4O WLAN Patient Wearable Monitor Product: 865352 Exchange part (service numbers): 453564615311 TELE PWM,802.lla/b/g,ECG only, US only 453564615331 TELE PWM,802.1 1 a/b/g,ECG&Sp02, US only Product Usage: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals. The MX4O is to be used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network. When the MX4O is connected the IntelliVue Information Center the IntelliVue Information Center provides the primary patient monitoring and alarming function. The MX4O does not automatically provide local monitoring or alarming when connected to the Information Center.

Z-1707-2017
Recall number
Z-1707-2017
Initiated
March 06, 2017
Classification
Class II
Status
Terminated
Quantity
2648

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Philips lntelliVue MX4O WLAN Patient Wearable Monitor may not automatically switch to Monitor Mode with audible alarms when association with central monitoring system is unsuccessful [incomplete]

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Philips lntelliVue MX4O WLAN Patient Wearable Monitor may not automatically switch to Monitor Mode with audible alarms when association with central monitoring system is unsuccessful [incomplete]

Code information

The following product number and Exchange part numbers containing the combination of HW Revision C.01 .01 with SW Revisions B.05 or B.06 of the Philips IntelliVue MX4O Patient Wearable Monitor

Distribution pattern

Worldwide Distribution - US Nationwide including the countries of: Australia Belgium Canada Denmark Finland France Germany Germany India Italy Netherlands Norway Qatar Singapore South Africa Sweden Switzerland United Kingdom