2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 23, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Medical Systems (Cleveland) Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Brilliance BigBore Oncology CT - 728243 Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories. This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures.

Z-1492-2017
Recall number
Z-1492-2017
Initiated
February 23, 2017
Classification
Class II
Status
Terminated
Quantity
11

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
When the user performs a retrospective respiratory gated 4D CT scan with the Bellows pulmonary gating device, the CT images might be reconstructed at a single phase, while the annotations on the images incorrectly indicate that the reconstruction is from the requested specific phases (0% to 90%).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

When the user performs a retrospective respiratory gated 4D CT scan with the Bellows pulmonary gating device, the CT images might be reconstructed at a single phase, while the annotations on the images incorrectly indicate that the reconstruction is from the requested specific phases (0% to 90%).

Code information

System Serial Number - 76002, 76008, 76010, 76011, 76016, 76020, 76021, 76022, 76023, 76024, 76025

Distribution pattern

Worldwide Distribution - US to AR, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MD, MI, MO, NC, NE, NJ, NM, NY, OH, TX & WI; Internationally to the following countries: Australia, Austria, Canada, China, Czech Republic, Germany, India, Israel, Jordan, Korea, Mexico, New Cldonia, Romania, Russian Federation, Turkey, & United Kingdom.

device · product 2 of 2

Brilliance BigBore Radiology CT - 728244 Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories. This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures.

Z-1493-2017
Recall number
Z-1493-2017
Initiated
February 23, 2017
Classification
Class II
Status
Terminated
Quantity
67

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
issue on Brilliance BigBore Oncology/Radiology CT with 4.2.0 software

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Philips identified a 4D CT pulmonary phase issue on Brilliance BigBore Oncology/Radiology CT with 4.2.0 software version through customer complaints. When the user performs a retrospective respiratory gated 4D CT scan with the Bellows pulmonary gating device, the CT images might be reconstructed at a single phase, while the annotations on the images incorrectly indicate that the reconstruction is from the requested specific phases (0% to 90%).

Code information

System Serial Number - 76001, 76005, 76007, 76009, 76013, 76014, 76017, 76018, 76019, System Serial Number - Unique Device Identifier 76058 (01)00884838059450(21)76058 76033 (01)00884838059450(21)76033 76048 (01)00884838059450(21)76048 76059 (01)00884838059450(21)76059 76015 (01)00884838059450(21)76015 76063 (01)00884838059450(21)76063 76066 (01)00884838059450(21)76066 76049 (01)00884838059450(21)76049 76040 (01)00884838059450(21)76040 76032 (01)00884838059450(21)76032 76061 (01)00884838059450(21)76061 76031 (01)00884838059450(21)76031 76006 (01)00884838059450(21)76006 76041 (01)00884838059450(21)76041 76035 (01)00884838059450(21)76035 76038 (01)00884838059450(21)76038 76065 (01)00884838059450(21)76065 76064 (01)00884838059450(21)76064 76053 (01)00884838059450(21)76053 76050 (01)00884838059450(21)76050 76060 (01)00884838059450(21)76060 76036 (01)00884838059450(21)76036 76056 (01)00884838059450(21)76056 76055 (01)00884838059450(21)76055 76037 (01)00884838059450(21)76037 76051 (01)00884838059450(21)76051 76044 (01)00884838059450(21)76044 76039 (01)00884838059450(21)76039 76012 (01)00884838059450(21)76012 76026 (01)00884838059450(21)76026 760016 (01)00884838059450(21)760016 760001 (01)00884838059450(21)760001 760005 (01)00884838059450(21)760005 760004 (01)00884838059450(21)760004 760003 (01)00884838059450(21)760003 760007 (01)00884838059450(21)760007 76072 (01)00884838059450(21)76072 76062 (01)00884838059450(21)76062 76043 (01)00884838059450(21)76043 760013 (01)00884838059450(21)760013 760014 (01)00884838059450(21)760014 76029 (01)00884838059450(21)76029 760010 (01)00884838059450(21)760010 76027 (01)00884838059450(21)76027 760011 (01)00884838059450(21)760011 760009 (01)00884838059450(21)760009 760006 (01)00884838059450(21)760006 76045 (01)00884838059450(21)76045 76054 (01)00884838059450(21)76054 76069 (01)00884838059450(21)76069 76034 (01)00884838059450(21)76034 76030 (01)00884838059450(21)76030 76047 (01)00884838059450(21)76047 76057 (01)00884838059450(21)76057 76028 (01)00884838059450(21)76028 76068 (01)00884838059450(21)76068 760012 (01)00884838059450(21)760012 760002 (01)00884838059450(21)760002

Distribution pattern

Worldwide Distribution - US to AR, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MD, MI, MO, NC, NE, NJ, NM, NY, OH, TX & WI; Internationally to the following countries: Australia, Austria, Canada, China, Czech Republic, Germany, India, Israel, Jordan, Korea, Mexico, New Cldonia, Romania, Russian Federation, Turkey, & United Kingdom.