openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
Bedside monitor CSM-1901(Life Scope G9) with main unit CU-192RA. The Problem only affects CSM-1901(Life Scope G9) communicating with CNS-6201A (PU-621RA) and CNS-9701A (MU-971RA).
The Pause function on central monitors will not automatically resume when connected to a Life Scope G9 patient monitor.
Code information
Lot/Unit Codes: G9 Serial Number: 230 (UDI: 01049319211121682100230) Serial Number: 232 (UDI: 01049319211121682100232) Life Scope G9 with version 01-19, 01-20, and 01-21 communicating with CNS-6201A (PU-621RA) with version 01-06, 02-10, or 02-40 software. Life Scope G9 with version 01-19, 01-20, and 01-21 communicating with CNS-9701A (MU-971RA) with version 01-89 or lower.
Distribution pattern
Worldwide - US Nationwide Distribution including the countries of Germany, Spain, Brazil, Paraguay, Bolivia, Chile, Colombia, Costa Rica, Mexico, Peru, Pakistan, Israel, Qatar, Saudi Arabia, South Africa, Tunisia, United Arab Emirates, Korea, Malaysia, Singapore, Thailand