Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 76723

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 13, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Nihon Kohden America Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Bedside monitor CSM-1901(Life Scope G9) with main unit CU-192RA. The Problem only affects CSM-1901(Life Scope G9) communicating with CNS-6201A (PU-621RA) and CNS-9701A (MU-971RA).

Z-1768-2017
Recall number
Z-1768-2017
Initiated
March 13, 2017
Classification
Class II
Status
Terminated
Recalling firm
Nihon Kohden America Inc
Quantity
654 units total (230 units in US)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The Pause function on central monitors will not automatically resume when connected to a Life Scope G9 patient monitor.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The Pause function on central monitors will not automatically resume when connected to a Life Scope G9 patient monitor.

Code information

Lot/Unit Codes: G9 Serial Number: 230 (UDI: 01049319211121682100230) Serial Number: 232 (UDI: 01049319211121682100232) Life Scope G9 with version 01-19, 01-20, and 01-21 communicating with CNS-6201A (PU-621RA) with version 01-06, 02-10, or 02-40 software. Life Scope G9 with version 01-19, 01-20, and 01-21 communicating with CNS-9701A (MU-971RA) with version 01-89 or lower.

Distribution pattern

Worldwide - US Nationwide Distribution including the countries of Germany, Spain, Brazil, Paraguay, Bolivia, Chile, Colombia, Costa Rica, Mexico, Peru, Pakistan, Israel, Qatar, Saudi Arabia, South Africa, Tunisia, United Arab Emirates, Korea, Malaysia, Singapore, Thailand