Event 76752

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 08, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Ortho-Clinical Diagnostics

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

enGen (TM) Laboratory Automation Systems using all TCAutomation(TM) Software Versions with the InOut Communication Interface, IVD

Z-2077-2017
Recall number
Z-2077-2017
Initiated
March 08, 2017
Classification
Class II
Status
Terminated
Quantity
US: 33 units; OUS: 43 units

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
Software anomaly

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Software anomaly; Thermo-Fisher Scientific initially discovered and Ortho-Clinical Diagnostics, subsequently, confirmed a software anomaly that may potentially result in miss-associated sample IDs involving the communication between enGen TCAutomation (TCA) Bypass modules (manufactured by Thermo-Fisher Scientific) and the VITROS Analyzers. To date, no occurrences of this issue have been observed on Orthos enGen(TM) Systems. Sample IDs may become miss-associated with the actual sample being processed due to discrepant information passed to the analyzer through InOut communication device, triggered by a specific set of events occurring between the Bypass module and the analyzer. --- All of the following conditions must be present for the issue to potentially occur: (1.) The Bypass module is in the process of positioning Sample A at the aspiration position, and (2.) At least two additional sample tubes (Samples B and C) are waiting to be aspirated in the Bypass module queue, and (3.) A Radio Frequency Identification (RFID) Tag Read failure occurs on the Sample A carrier, and (4.) The VITROS System is not available to aspirate on the track (for example, the analyzer is initializing, or the user is making entries on the diagnostic screen, or samples are being aspirated from the front of the analyzer, etc.). --- If the software anomaly occurs and the issue is present, the Sample ID in the INOUT communication device software queue becomes asynchronous with the actual Sample ID physical queue, and the VITROS assay result will be reported for the wrong Sample ID. All subsequent samples will have miss-associated results until the Bypass module is reset or the queue is empty. --- This issue could potentially miss-associate a sample/patient ID with a test result of a different sample. Depending on the test involved, patients health condition, etc., the potential health risk posed to the affected patient could range from no risk to serious risk, and potentially could lead to misdiagnosis or inappropriate medical intervention. To date, no patient harm was reported related to this product issue.

Code information

US Serial Numbers/J Numbers: B3ZHP2J/JB3ZHP2J; 736BM4J/J736BM4J; 953289/J953289; 953252/J953252; HJ5K84J/JHJ5K84J; FCM073J/JFCM073J; HPKG25J/JHPKG25J; GWXQ52J/JGWXQ52J; GBX395J/GBX395J; JFBX395J/JFBX395J; FXCBG3J/JFXCBG3J; 3RHC23J/J3RHC23J; JG7BFB5J/JG7BFB5J; 953270/J953270; 953221/J953221; 54FC01J/J54FC01J; 953209/J953209; 5J4B45J/J5J4B45J; 5LQ6Q2J/ J5LQ6Q2J; 2Z3YV1J/J2Z3YV1J; FRKMB2J/JFRKMB2J; 5GV6W2J/5GV6W2J; 7KSC15J/J7KSC15J; 953276/J953276; J2QKG25J/J2QKG25J; DK2BG3J/JDK2BG3J; 953266/J953266; J86PWC5J/ J86PWC5J; JXBQW4J/JJXBQW4J; CXL941J/JCXL941J; J953224/J953224; 36PWC5J/J36PWC5J; 4HX7S4J/J4HX7S4J --- Foreign Serial Numbers/J #s: CRKM82/J31265; 2S5OP0J/J31067; C4Z655J/ J31421; 6VR7K4J/J31401; 90JYS22/J31350; HWJZY4J/JHWJZY4J; J3J4B45J/J3J4B45J; JCT20Q3J/JCT20Q3J; J953234/J953234; J41BCY4J/J41BCY4J; J953244/J953244; J1HX7S4J/ J1HX7S4J; J16XQH4J/J16XQH4J; GWJZY4J/ GWJZY4J; DWXO52J/DWXO52J; 17QP94J/J20022; BGV6W2J/ J20003; H5PWC5J/J271371; 2LSSF5J/J271377; 1K5K84J/ J271264; 8W7WG3J/J85206; 96XQHAJ/ J7023645; 1LSSF5J/J7025050; 86T205J/ J7024893; C3JTP4J/J7023847; D3JTP4J/J7023874; C1JNZ5J/J7023730; 7DRJ44J/J7023728; 7YCF43J/J7023435; 18BFB5J/J7025017; J953222/J953222; J2V6LW4J/J2V6LW4J; J56T205J/J56T205J; BGV6W2J/J57663; 4drj44j/J800103; 6KSC15J/J1801116; FWGFC4J/J75663; 7GV6W2J/J75615; BT9PB5J/ J75666; JT6LW4J/ J30372; 38769917923/JTAINAN1; F2250767/J3055503; F1150698/ J3055496.

Distribution pattern

MA, NJ, NY, NY, PA,DE, MD, WV, NC, GA, FL, TN, MS, OH, MI, IL, MO, NE, AR, TX, CO, ID, TX, CA and Foreign distribution to the following countries: Belgium, Bermuda, Brazil, Canada, Chile, Denmark, England, France, Italy, Mexico, Norway, Portugal, Romania, Singapore, Spain, Sweden, Taiwan, United Kingdom.