openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
The PreludeSYNC Radial Compression Device is a sterile, single use disposable device used to assist in gaining and maintaining hemostasis of the radial and ulnar artery following catheterization procedures. It has a soft wristband with a secure hook and loop fastener. The band delivers adjustable compression of the puncture site with an inflatable bulb, and a check valve for easy inflation and deflation with a syringe. A clear curved backer plate provides optimal visualization of the puncture site and ease of placement.
Merit Medical Systems, Inc. announces a voluntary field action for the PreludeSYNC Radial Compression Device because particulate generated during the manufacturing process may enter the valve, preventing a complete seal of the inflation valve.
Code information
Catalog Number: SRB24AC, Lot Numbers: H1068590S1, H1078177, H1078180, H1083628; Catalog Number: SRB24MED, Lot Numbers: H1068587S1,H1073974, H1078179, H1078181, H1083629, H1083630; Catalog Number: SRB29AC, Lot Numbers: H1068591S1, H1078183, H1083631; Catalog Number: SRB29MED, Lot Numbers: H1068589S1, H1078184, H1078185, H1083633.
Distribution pattern
Nationwide, Canadian and other foreign consignees. No VA/govt/military consignees.