device · product 1 of 6
EXACTECH Equinoxe Reverse Starter Reamer - Catalog #: 321-05-01
- Recall number
- Z-2663-2017
- Initiated
- May 25, 2017
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Exactech, Inc.
- Quantity
- 382
App-derived interpretation
Instructions for use (IFU) is updated to clarify the proper technique and instrumentation to ream the Glenoid, i.e., to include a caution statement that the surgeon should ream on-axis and that reaming off-axis may result in instrument breakage.
Official device-enrichment evidence · Sourced
Error in labeling
Inspect official wording and provenance
Reason for recall
Instructions for use (IFU) is updated to clarify the proper technique and instrumentation to ream the Glenoid, i.e., to include a caution statement that the surgeon should ream on-axis and that reaming off-axis may result in instrument breakage.
Code information
All Lots
Distribution pattern
AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico Australia, Austria, Brazil, Canada, China, France, Germany, Greece, India, Italy, Korea, Luxembourg, Singapore, South Africa, Spain, Sweden, Switzerland, The Netherlands and United Kingdom