Event 77260

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 28, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Exact Sciences Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Exact Sciences Sample Mixer 2. For In Vitro Diagnostic Use. The Exact Sciences Sample Mixer 2 is intended for mixing stool samples in a Sample Container as part of pre-analytic sample processing for in vitro diagnostic tests

Z-2219-2017
Recall number
Z-2219-2017
Initiated
April 28, 2017
Classification
Class II
Status
Terminated
Quantity
15

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Exact Sciences is correcting all Sample Mixer v2 Upgrade units because of incidents of Sample Mixer door damage, which has occurred when the Sample Holder opens during a run and the Sample Holder lid makes contact with the Sample Mixer door.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Exact Sciences is correcting all Sample Mixer v2 Upgrade units because of incidents of Sample Mixer door damage, which has occurred when the Sample Holder opens during a run and the Sample Holder lid makes contact with the Sample Mixer door.

Code information

Equipment ID / Serial Number: MIX-0051 / MX00002051, MIX-0052 / MX00002032, MIX-0053 / MX00002028, MIX-0054 / MX00002052, MIX-0055 / MX00002043, MIX-0056 / MX00002040, MIX-0057 / MX00002050, MIX-0058 / MX00002036, MIX-0059 / MX00002027, MIX-0060 / MX00002015, MIX-0061 / MX00002016, MIX-0063 / MX00002039, MIX-0064 / MX00002033, MIX-0065 / MX00002053, MIX-0066 / MX00002014,

Distribution pattern

US to WI only