device · product 1 of 1
Berchtold Chromophare F300 Exam Light CHROMOPHARE examination lights are medical lights used for local illumination of the patients' body so that illnesses, injuries and disabilities can be diagnosed and treated
- Recall number
- Z-2248-2017
- Initiated
- April 21, 2017
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Stryker Communications
- Quantity
- 57 units
App-derived interpretation
The exam light may have a tolerance issue with the adapter assembly, which could potentially not allow the snap ring to be seated correctly. If this is the case, there could be insufficient mount force that may cause the equipment to fall, resulting in serious injury.
Official device-enrichment evidence · Sourced
Component design/selection
Inspect official wording and provenance
Reason for recall
The exam light may have a tolerance issue with the adapter assembly, which could potentially not allow the snap ring to be seated correctly. If this is the case, there could be insufficient mount force that may cause the equipment to fall, resulting in serious injury.
Code information
Serial Numbers: 7735170-X68377, 7735170-X68397, 7735170-X14846, 7735170-X15688, 7735170-X14492, 7735170-X14495, 7735170-X14497, 7735170-X14501, 7735170-X14502, 7735170-X14504, 7735170-X14506, 7735170-X14507, 7735170-X14508, 7735170-X14510, 7735170-X68383, 7735170-X68384, 7735170-X68385, 7735170-X68386, 7735170-X68393, 7735170-X68394, 7735170-X68395, 7735170-X68396, 7735170-X68392, 7735170-X68400, 7735170-X68401, 7735170-X15607, 7735170-X15608, 7735170-X15609, 7735170-X15610, 7735170-X15611, 7735170-X15612, 7735170-X15613, 7735170-X15614, 7735170-X15615
Distribution pattern
Worldwide Distribution - US and Canada