device · product 1 of 1
Beckman Coulter Synchron AQUA CAL Level 1, Part No. 471288 Product Usage: Used for the system calibration of Na, K, Cl, Urea N, Urea, Glucose Creatine, Ca, CO2, and P in UniCel DxC Systems.
- Recall number
- Z-2730-2017
- Initiated
- May 15, 2017
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Beckman Coulter Inc
- Quantity
- 4444 AQUA CAL Level 1 kits
App-derived interpretation
The firm identified a negative shift of 5mmol/L for sodium recovery in Synchron Control Level 3 when calibrated with AQUA Cal 1.
Official device-enrichment evidence · Sourced
Nonconforming Material/Component
Inspect official wording and provenance
Reason for recall
The firm identified a negative shift of 5mmol/L for sodium recovery in Synchron Control Level 3 when calibrated with AQUA Cal 1.
Code information
Lot No. 72645074, Exp. Date 30 Apr 2018
Distribution pattern
US Nationwide Distribution in the states of AL, AR, AK, CA, CO, DE, FL, GA, ID, IL, IN, IO, KS, KY, LA, MA, MD, MS, MI, ME, MO, NE, NV, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WV, WI, and Puerto Rico Brunei, Canada, China, Czech Republic, France, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Lebanon, Mexico, Netherlands, Panama, Paraguay, Philippines, Qatar, South Africa, Taiwan, Thailand, Turkey, United Arab Emirates, & United Kingdom.