Event 77384

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 24, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
NeuroLogica Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Samsung GU60A system- Digital X-ray Imaging Systems are intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications

Z-3036-2017
Recall number
Z-3036-2017
Initiated
May 24, 2017
Classification
Class II
Status
Terminated
Recalling firm
NeuroLogica Corporation
Quantity
87 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
When performing the Stitching function on GU60A system, unintended area is exposed to radiation

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

When performing the Stitching function on GU60A system, unintended area is exposed to radiation

Code information

Serial Numbers: S03PM3ADA00001T S03PM3ADC00001H S03PM3ADC00002J S03PM3AF200004X S03PM3AF200005N S03PM3AF200006E S03PM3AF300001K S03PM3AF300002Z S03PM3AF300007X S08HM3AF300001J S08HM3AF300002B S08HM3AF300004H S08HM3EF400001W S08HM3EF400002A S08HM3DF400002Y S08HM3EF400003H S08HM3EF400004K S08HM3DF400003R S08HM3DF400004B S08HM3EF400005E S08HM3DF500001B S08HM3CF500001T S08HM3EF500001K S08HM3DF500002F S08HM3BF500001L S08HM3BF500002N S08HM3CF600001R S08HM3CF600003V S08HM3DF600002P S08HM3DF600001H S08HM3DF600003L S08HM3DF600004F S08HM3DF600005Y S08HM3DF600006W 508UM3BF700001L S08HM3AFB00001Y 508UM3BG100002K 508UM3BG100001T 50FFM3AG200001V 50FFM3AG200002K 50FFM3AG200003P 50FFM3AG200004E 50FFM3AG200005J 50FFM3AG300001X 50FFM3BG600003E 50FFM3BG600002V 50FFM3BG600001A 50FFM3CG700004E 50FFM3CG700002B 50FFM3CG700003M 50FFM3CG700001P 50FFM3CG700006T 50FFM3CG700007J 50FFM3CG700005A 50FFM3CG700008N 50FFM3AG700002L 50FFM3AG700001Y 50FFM3AG800004K

Distribution pattern

Nationwide Distribution