Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 77456

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 22, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Medical Solutions USA, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Syngo.plaza, Picture Archiving and Communications System(PACS), Model 10863171 Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific work flows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.

Z-0168-2018
Recall number
Z-0168-2018
Initiated
May 22, 2017
Classification
Class II
Status
Terminated
Quantity
2 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Siemens is releasing a non-medical software application LTA Incomplete Archive Check Tool V1.0. It is intended to eliminate an issue that occurs during syngo.plaza de-archiving from Dicom LTA. It has been defined that the number of de-archived images is less than the count of the archived images for the series. The issues were discovered through Siemens Global Complaint Handling System.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Siemens is releasing a non-medical software application LTA Incomplete Archive Check Tool V1.0. It is intended to eliminate an issue that occurs during syngo.plaza de-archiving from Dicom LTA. It has been defined that the number of de-archived images is less than the count of the archived images for the series. The issues were discovered through Siemens Global Complaint Handling System.

Code information

non-medical software application Check Tool V1 .0 (L TA Incomplete Archive Check Tool V1 .0)

Distribution pattern

Distribution to MN