Exact FDA event identifier

Event 77515

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 05, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
CSL Behring GmbH

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

SIEMENS N Protein Standard SL, for use on the BN II, BN ProSpec Systems, REF/Product Code OQIM 13, Siemens Material Number (SMN)10446073

Z-0190-2018
Recall number
Z-0190-2018
Initiated
June 05, 2017
Classification
Class II
Status
Terminated
Recalling firm
CSL Behring GmbH
Quantity
N/A

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Siemens Healthcare Diagnostics has confirmed a negative bias of approximately -20% for results obtained with N Antiserum to Human Ceruloplasmin when calibrating Ceruloplasmin with various lots of N Protein Standard SL. N/T Protein Controls SL for Ceruloplasmin are similarly biased. --- The affected standards and controls for the affected lots were incorrectly value assigned, which led to a lower result compared to correctly assigned standards. Note: N Antiserum to Human Ceruloplasmin (REF OUIE) works as intended. --- Siemens investigation indicated that N Protein Standard SL and corresponding N/T Protein Control SL (Levels Low, Mid and High) are inaccurately assigned for Ceruloplasmin.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Siemens Healthcare Diagnostics has confirmed a negative bias of approximately -20% for results obtained with N Antiserum to Human Ceruloplasmin when calibrating Ceruloplasmin with various lots of N Protein Standard SL. N/T Protein Controls SL for Ceruloplasmin are similarly biased. --- The affected standards and controls for the affected lots were incorrectly value assigned, which led to a lower result compared to correctly assigned standards. Note: N Antiserum to Human Ceruloplasmin (REF OUIE) works as intended. --- Siemens investigation indicated that N Protein Standard SL and corresponding N/T Protein Control SL (Levels Low, Mid and High) are inaccurately assigned for Ceruloplasmin.

Code information

Lot Numbers(UDI): 083611A (00842768007590083611A20170711), 083611B (00842768007590083611B20170711), 083611C (00842768007590083611C20170711), 083611D (00842768007590083611D20170711), exp. date 2017-07-11; 083612 (0084276800759008361220171014), 083612A (00842768007590083612A20171014), 083612B (00842768007590083612B20171014), exp. date 2017-10-14; 083614A (00842768007590083614A20180127), 083614B (00842768007590083614B20180127), 083614C (00842768007590083614C20180127), exp. date 2018-01-27; 083615 (0084276800759008361520180411), 083615A (00842768007590083615A20180411), 083615B (00842768007590083615B20180411), 083615C (00842768007590083615C20180411), exp. date 2018-04-11; 083616B (00842768007590083616B20180730), 083616C (00842768007590083616C20180730), 083616D (00842768007590083616D20180730), exp. date 2018-07-30.

Distribution pattern

Worldwide including the countries of United States, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia-Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Cyprus, Denmark, Egypt, Finland, France, Germany, Greece, Great Britain, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, japan, Kazakhstan, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, Mexico, Morocco, Netherlands, New Zealand, Norway, Poland, Portugal, Republic of Korea (S), Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, South Africa, Switzerland, Syria, Taiwan, Thailand, Trinidad-Tobago, Turkey, Turkmenistan, United Arab Emirates, Uruguay, Vatican, Venezuela,

device · product 2 of 2

SIEMENS N/T Protein Control SL for use on the BN II, BN ProSpec Systems: N/T Protein Control SL/L: REF/Product Code OQIN19, SMN 10446079 (Low); N/T Protein Control SL/M, REF/Product Code OQIO19, SMN 10446085 (Mid); N/T Protein Control SL/H, REF/Product Code OQIP19, SMN 10446089 (High);

Z-0191-2018
Recall number
Z-0191-2018
Initiated
June 05, 2017
Classification
Class II
Status
Terminated
Recalling firm
CSL Behring GmbH
Quantity
N/A

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Siemens Healthcare Diagnostics has confirmed a negative bias of approximately -20% for results obtained with N Antiserum to Human Ceruloplasmin when calibrating Ceruloplasmin with various lots of N Protein Standard SL. N/T Protein Controls SL for Ceruloplasmin are similarly biased. --- The affected standards and controls for the affected lots were incorrectly value assigned, which led to a lower result compared to correctly assigned standards. Note: N Antiserum to Human Ceruloplasmin (REF OUIE) works as intended. --- Siemens investigation indicated that N Protein Standard SL and corresponding N/T Protein Control SL (Levels Low, Mid and High) are inaccurately assigned for Ceruloplasmin.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Siemens Healthcare Diagnostics has confirmed a negative bias of approximately -20% for results obtained with N Antiserum to Human Ceruloplasmin when calibrating Ceruloplasmin with various lots of N Protein Standard SL. N/T Protein Controls SL for Ceruloplasmin are similarly biased. --- The affected standards and controls for the affected lots were incorrectly value assigned, which led to a lower result compared to correctly assigned standards. Note: N Antiserum to Human Ceruloplasmin (REF OUIE) works as intended. --- Siemens investigation indicated that N Protein Standard SL and corresponding N/T Protein Control SL (Levels Low, Mid and High) are inaccurately assigned for Ceruloplasmin.

Code information

SMN 10446079: Lot (UDI): 084647A (00842768016127084647A20170913), 084647C (00842768016127084647C20170913) exp. date 2017-09-13; 084648(0084276801612708464820180430), 084648C (00842768016127084648C20180430), 084648H (00842768016127084648H20180430), exp. date 2018-04-30. --- SMN 10446085: Lot (UDI): 084749D (00842768016134084749D20170917) exp. date 2017-09-17; 084750 (03 0084276801613408475020180203), 084750B (00842768016134084750B20180203), 084750E (00842768016134084750E20180203) exp. date 2018-02-03. --- SMN 10446089: Lot (UDI) 084847A (00842768016110084847A20170613), 084847B (00842768016110084847B20170613) exp. date 2017-06-13; 084848A (00842768016110084848A20171217), 084848D (00842768016110084848D20171217), 084848E (00842768016110084848E20171217), 084848H (00842768016110084848H20171217), 084848L (00842768016110084848L20171217) exp. date 2017-12-17; 084849 (0084276801611008484920180704), 084849C (00842768016110084849C20180704), 084849D (00842768016110084849D20180704) 084849E (00842768016110084849E20180704) exp. date 2018-07-04.

Distribution pattern

Worldwide including the countries of United States, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia-Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Cyprus, Denmark, Egypt, Finland, France, Germany, Greece, Great Britain, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, japan, Kazakhstan, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, Mexico, Morocco, Netherlands, New Zealand, Norway, Poland, Portugal, Republic of Korea (S), Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, South Africa, Switzerland, Syria, Taiwan, Thailand, Trinidad-Tobago, Turkey, Turkmenistan, United Arab Emirates, Uruguay, Vatican, Venezuela,