Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 77697

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 27, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Hologic, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Tomcat Instrument, REF ASY-07379, Hologic Incorporated, automated laboratory sample processing instrument.

Z-2864-2017
Recall number
Z-2864-2017
Initiated
June 27, 2017
Classification
Class II
Status
Terminated
Recalling firm
Hologic, Inc
Quantity
164 units

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software anomaly

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A Tomcat Instrument software anomaly that causes the pipettor arm to make physical contact with the mucous shelf which could potentially lead to incorrect patient results.

Code information

TC00000050,TC00000058,TC00000059,TC00000060,TC00000061,TC00000062,TC00000063,TC00000064,TC00000065,TC00000067,TC00000068,TC00000069,TC00000070,TC00000071,TC00000072,TC00000073,TC00000074,TC00000075,TC00000076,TC00000077,TC00000078,TC00000079TC00000080,TC00000081,TC00000082,TC00000083,TC00000085,TC00000086,TC00000087,TC00000088,TC00000089,TC00000090,TC00000091,TC00000092,TC00000093,TC00000094,TC00000096,TC00000097,TC00000098,TC00000099,TC00000100,TC00000101,TC00000102,TC00000103,TC00000104,TC00000105,TC00000106,TC00000107,TC00000108,TC00000109TC00000110,TC00000111,TC00000112,TC00000113,TC00000114,TC00000115,TC00000116,TC00000117,TC00000118,TC00000119,TC00000120,TC00000121,TC00000123,TC00000124,TC00000125,TC00000126,TC00000127,TC00000128,TC00000129,TC00000130,TC00000131,TC00000132,TC00000133,TC00000134,TC00000136,TC00000137,TC00000138,TC00000139,TC00000140,TC00000141,TC00000142,TC00000143,TC00000144,TC00000145,TC00000146,TC00000147,TC00000148,TC00000149,TC00000150,TC00000151,TC00000152,TC00000153,TC00000154,TC00000155,TC00000156,TC00000157,TC00000158,TC00000159,TC00000160,TC00000161,TC00000162,TC00000163,TC00000164,TC00000165,TC00000166,TC00000167,TC00000168,TC00000170,TC00000171,TC00000172,TC00000173,TC00000174,TC00000175,TC00000177,TC00000178,TC00000179,TC00000181,TC00000182,TC00000183,TC00000184,TC00000185,TC00000187,TC00000188,TC00000189,TC00000190,TC00000191,TC00000192,TC00000193,TC00000195,TC00000197,TC00000198,TC00000199,TC00000200,TC00000201,TC00000202,TC00000203,TC00000204,TC00000206,TC00000207,TC00000208,TC00000209,TC00000210,TC00000211,TC00000212,TC00000213,TC00000214,TC00000215,TC00000227,TC00000236,TC00000242,TC00000249,TC00000254,TC00000263,TC00000264,TC00000266,TC00000270,TC00000273,TC00000274,TC00000280,TC00000283,TC00000284,TC00000285,TC00000289,TC00000291

Distribution pattern

Worldwide distribution. US Nationwide and countries: BE, CA, CH, ES, FR, GB, IE, IT, JP, LU, NL, SE, SK, and TR.