Event 77711

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 05, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Baxter Healthcare Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

HomeChoice Automated PD system Model numbers: 5C4471, 5C4471R Product Usage: The HomeChoice Automated Personal Cycler is intended for automatic control of dialysate solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

Z-3040-2017
Recall number
Z-3040-2017
Initiated
July 05, 2017
Classification
Class II
Status
Terminated
Quantity
15144 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their Peritoneal Dialysis (PD) therapy, damaging the cassettes for the HomeChoice or HomeChoice PRO cyclers. If damaged cassettes are used, the cyclers may not consistently detect very small holes/cuts in the sheeting of the cassette in the patient valve region and the cycler may deliver air into the patient.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their Peritoneal Dialysis (PD) therapy, damaging the cassettes for the HomeChoice or HomeChoice PRO cyclers. If damaged cassettes are used, the cyclers may not consistently detect very small holes/cuts in the sheeting of the cassette in the patient valve region and the cycler may deliver air into the patient.

Code information

all codes

Distribution pattern

Worldwide Distribution

device · product 2 of 2

HomeChoice Pro Automated PD system Model numbers: 5C8310, 5C8310R Product Usage: The HomeChoice Automated Personal Cycler is intended for automatic control of dialysate solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

Z-3041-2017
Recall number
Z-3041-2017
Initiated
July 05, 2017
Classification
Class II
Status
Terminated
Quantity
23732 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their Peritoneal Dialysis (PD) therapy, damaging the cassettes for the HomeChoice or HomeChoice PRO cyclers. If damaged cassettes are used, the cyclers may not consistently detect very small holes/cuts in the sheeting of the cassette in the patient valve region and the cycler may deliver air into the patient.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their Peritoneal Dialysis (PD) therapy, damaging the cassettes for the HomeChoice or HomeChoice PRO cyclers. If damaged cassettes are used, the cyclers may not consistently detect very small holes/cuts in the sheeting of the cassette in the patient valve region and the cycler may deliver air into the patient.

Code information

all codes

Distribution pattern

Worldwide Distribution