device · product 1 of 1
AQUIOS CL Flow Cytometer System The AQUIOS CL Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping.
- Recall number
- Z-3135-2017
- Initiated
- August 21, 2017
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 213 units (34 units in the US)
App-derived interpretation
incorrect setting using a proprietary software
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
Beckman Coulter has identified that an incorrect setting using a proprietary software configuration tool can result in the Sample Prep Probe of the AQUIOS CL drawing aspirant from the wrong tube in a cassette which manifests in incorrect values being reported out of the laboratory for some patient samples.
Code information
All software versions.; Serial numbers - BA30048, AY20032, BA01007, AZ19046, AZ12030, BA25046, BA05011, AY51072, BA01002, AZ15044, BA30049, BA30050, BA05015, AZ19047, BA18029, AY47071, AZ15038, BA23036, BA23035, AZ23058, AZ20057, AY47063, BA15024, BA15023, AZ45080, BA01010, AZ25070, AZ32071, AZ12033, AZ25065, AZ25068, AY51075, AZ19052, BA15025
Distribution pattern
Worldwide Distribution: United States (nationwide) and countries of: Angola, Australia, Botswana, Canada, Chile, Columbia, Ghana, Israel, Malaysia, Namibia, Pakistan, Panama, South Africa, Uganda, France, Switzerland, Poland, Portugal, Germany, Belgium, Finland, Czech Republic, Denmark, Netherland, Turkey, Spain, Great Britain, Romania, Brazil, Italy, Sweden, Ghana, Colombia, Lesotho and Zambia